INPUT PARAMETER SETTINGS
8
Input Parameter Settings
You might want to change the default configuration of the digital inputs, like the trigger
level, termination, or rising/falling edge selection of the input channels. Depending on
your device (with or without hardware extension) not all options might be available.
Table
gives an overview.
Table 4:
Digital input configuring.
Available features of the different models.
quTAU
quTAU (H)
trigger level
fixed
adjustable per channel
50 Ω termination on/off
globally
per channel
rising/falling edge selection
-
per channel
channel delay
-
per channel
•
Trigger level: necessary threshold voltage to trigger an event.
•
Termination: switches the termination of the input channels between 50 Ω and 5 kΩ.
•
Edge selection: determines whether the event is triggered at the rising or the falling
edge of a signal exceeding the voltage threshold.
•
Channel delay: Sets additional delays for the input channels. Please note that this
currently doesn’t influence coincidence counting. It affects all histograms, though.
Please note that these settings are saved in the device and therefore persistent over the
different software realizations. A restart of the device, however, will reset all settings to
the default state.