QUICK2025 manual operation
Page 5
System Interface
PASSWORD
:
System password
SOUND
:
alarming switch of the host
(
ON---with alarm sound OFF—without alarm sound
)
ADDED IR
:
heating switch of the outside group of the bottom Infrared heating tubes
(
ON---switch is on
;
OFF---switch is off
)
COM BAUD
:
Default baud is19200
Data Interface
Data interface has data of the ten working modes which parameter can be changed according to demands.
Calibrating set as follows. Reference data in the column
“
CAL
”
means the parameters whether calibrating or not.
If “CAL” is on, the parameter has been calibrated and is saved in the process. Change “ON” to “OFF” if user
wants to cancel the calibration.
QUICK BGA 2025 has ten working modes which parameter can be changed according to demands. Calibrating
set as follows.
T0 S0 T1 S1 T2 S2 T3 S3 TL
TIR
TB1
TB2
BF
A
SEN
TYP
CA
L
0 90 40 145 50 160 55 200
30 183
150
140
160 9 0
0 0
1 90 55 145 50 155 40 200
20 183
165
130
150 9 0
1 0
2 90 40 140 50 150 50 200
15 183
160
125
145 9 0
0 0
3 90 45 140 60 155 40 200
20 183
160
135
150 9 0
1 0
4 80 55 145 65 155 45 200
25 183
170
140
160 9 0
0 0
5 80 60 165 45 175 45 230
25 217
170
150
170 9 0
1 0
6 90 60 160 45 170 40 230
20 217
170
160
180 9 0
0 0
7 90 60 165 40 175 45 230
25 217
170
160
180 9 0
1 0
8 90 60 160 45 175 40 230
20 217
170
170
180 9 0
0 0
9 90
60
165
40
170
40
230
20
217
170
150
170 9 0 1 0
Note
:
typ:0—sold, 1—deso sen:0—ir cal:0—no cal Time unit
:“
S
”
Temperature unit
:“
℃
”
6
、
Install and Connecting of the Equipment
6.1 check the package of the equipment
* Check the package of the equipment is good or not.
* Take out the QUICK BGA 2025 rework system from the package and place it on the solid working desk.
* Check every unit is good or not. If not, please replace the unit from our company or agents in time.