background image

Page 16    Hum Bug Reference Manual 
 

Quest Scientific 

C.

 

The  output  of  the  Hum Bug  has  too  much 

baseline  noise

  (random,  wide-band 

noise). 

1.

 

Are you attempting to use the Hum Bug on very low level signals (less than 1 mV)?  
If so, circuit noise from the Hum Bug may be detected on the output. 

 

Compare  the  input  and  output  on  two  channels  of  your  oscilloscope.    The 

baseline noise

 should look identical on both signals for levels down to 1 mV.  

Below this level additional circuit induced noise may appear on the output.  If 
the baseline noise on your input signal is below 1 mV and the additional noise 
introduced on the output is a problem in your application, it may be necessary 
to increase the gain of your amplifier. 

2.

 

The 

baseline  noise

  on  the  output  of  the  Hum Bug  is  greater  than  on  the  input,  even 

when its amplitude is greater than 1 mV. 

 

Directly  ground  the  input  connector  of  the  Hum Bug  and  observe  both  the 
input  and  output  at  the  same  time.    Is  the 

baseline  noise

  on  the  output  more 

than  a  few  mV  even  when  the  input  is  grounded?    If  so,  there  may  be  a 
problem with grounding.  Refer to section D1. 

3.

 

The baseline noise is only a few mV when the input is grounded but much larger than 
expected when an input signal is connected to the Hum Bug. 

 

Perhaps  the  grounding  of  the  Hum Bug  and  associated  equipment  is  the 
problem.  See section D1. 

 

Are  you  connecting  the  Hum Bug  to  an  unusual  piece  of  equipment?    If  so, 
check the output impedance of the device.  Although the Hum Bug will adapt 
to  a  wide  range  of  input  characteristics,  a  signal  source  with  a  very  high 
impedance could present a problem.  This problem can be solved by passing 
the signal through a suitable impedance-matching amplifier (buffer) before it 
enters the Hum Bug. 

D.

 

The Hum Bug adds 50/60 Hz noise to the output. 

1.

 

Incorrect grounding is the most likely cause of this problem. 

 

Confirm  that  the  grounding  prong  on  the  power  cable  is  connected  to  the 
building ground at your preferred grounding point.  Avoid using more than 1 
service  outlet  for  your  setup.    A  good  way  to  bring  all  of  the  instrument 
grounds in your system to one point is to use a power bar.  The ground on the 
power  bar  becomes  the  final  common  ground  connected  to  the  wall  outlet.  
However,  poor  ground  connections  are  a  frequent  problem  with  inexpensive 
power distribution bars. 

 

Make sure that the signal ground on your probe preamplifier is not attached to 
the  Hum Bug  other  than  through  the  BNC  cable  connecting  the  two 
instruments.  This helps prevent ground loops. 

Содержание Hum Bug

Страница 1: ...NTS Introduction 3 Theory Of Operation 4 Installation 6 Controls 7 Input Signal 7 Noise Generated By Monitors 8 Sources Of Complex Noise 9 Adaptation Rate 9 Front Panel 10 Back Panel 11 Trouble Shooting 12 Specifications 18 Warranty And Contacts 18 ...

Страница 2: ...Page 2 Hum Bug Reference Manual Quest Scientific ...

Страница 3: ...oise without any need for adjustment The Hum Bug is not a filter It is a new class of instrument capable of eliminating electrical interference from analogue signals without creating phase shifts frequency loss amplitude errors DC shifts time delays or digital distortion It effectively eliminates 50 60 Hz2 noise and harmonics3 without altering the frequency characteristics of the desired signal4 e...

Страница 4: ... 5 to 10 seconds under ideal conditions but is slower when low amplitude noise is dominated by ongoing physiological activity and or noise harmonics are greater than 1 kHz Therefore the noise replica may lag behind if the noise characteristics suddenly change Under these conditions some noise will transiently appear on the output while the Hum Bug adapts to the new characteristics The major advant...

Страница 5: ...Hum Bug Reference Manual Page 5 Quest Scientific ...

Страница 6: ... operating mode During your initial evaluation of the Hum Bug it is a good idea to turn off any computer monitors in the vicinity of your recording setup This will eliminate possible confusion between 50 60 Hz noise and any additional interference associated with your monitors See the section on monitor noise for more details After an initial warm up period of approximately thirty seconds the Hum ...

Страница 7: ... Pushing it will clear the noise replica and force the Hum Bug to generate a new noise replica from scratch In effect this causes the Hum Bug to immediately forget the noise waveform This function is seldom necessary since the Hum Bug automatically adapts to changing noise conditions INPUT SIGNAL The Hum Bug will operate on analogue signals originating from virtually any laboratory equipment inclu...

Страница 8: ...erference is the same as the power mains so it will be eliminated by the Hum Bug The second form of interference referred to as refresh noise is generated each time a monitor refreshes the image displayed on its screen The frequency of this interference will depend on the video mode assigned to the monitor In most cases it will differ from the power mains and will not be recognized as noise by the...

Страница 9: ...ncy response is acceptable will depend on your application Feel free to consult Quest Scientific for more information ADAPTATION RATE The Hum Bug adapts to changing noise characteristics in the input signal Under normal conditions these characteristics evolve slowly and the Hum Bug is able to maintain effective noise elimination However noise will appear in the output signal following a sudden cha...

Страница 10: ...lowly over time However touching a manipulator approaching the electrode or lifting the electrode from the experimental preparation can cause additional but transient noise on the recorded signal Using HOLD during these manipulations will keep the Hum Bug from chasing the transient noise As a result effective cancellation of ongoing noise can resume immediately after the manipulation is complete ...

Страница 11: ...Hum Bug Reference Manual Page 11 Quest Scientific BACK PANEL ...

Страница 12: ...t to the expected behavior described for each procedure Step One Set up your equipment in a manner similar to the illustration shown below The objective is to simultaneously view the input and output of the Hum Bug on different channels of an oscilloscope while recording a noisy signal This description assumes that you are recording neuronal activity using a micropipette but a similar procedure co...

Страница 13: ...ed as the noise replica grows to the required size When adaptation is complete the Indicator should be off Toggle BYPASS mode on red LED and off green LED to observe the signal with and without noise cancellation Compare the input and output signals You should find that the primary signal of interest is not modified by the Hum Bug Step Four Observe the Hum Bug adapt to changing noise conditions On...

Страница 14: ...us on different types of abnormal Hum Bug behavior The tasks outlined in each section are designed to help you identify the specific cause of abnormal behavior in your Hum Bug A The Hum Bug fails to operate 1 No LEDs are illuminated on the panel switches and the indicator does not illuminate when you push clear Check that the power switch is on Check that the power cable on the back panel is pushe...

Страница 15: ...ce Perhaps the peak to peak level of noise is too large for the Hum Bug to completely cancel If so the indicator will continue to flash red Connect the input signal of the Hum Bug to an oscilloscope and determine the noise amplitude The Hum Bug cannot cancel noise greater than 1 Volt peak to peak If your noise levels are higher than this limit you may have to reduce the gain in your preamplifier P...

Страница 16: ...put is grounded but much larger than expected when an input signal is connected to the Hum Bug Perhaps the grounding of the Hum Bug and associated equipment is the problem See section D1 Are you connecting the Hum Bug to an unusual piece of equipment If so check the output impedance of the device Although the Hum Bug will adapt to a wide range of input characteristics a signal source with a very h...

Страница 17: ...und wires will make inappropriate connections between grounds in your setup and the grounds in the power mains Check your cables to make sure that the braided shield is connected to the ground ring of the BNC connector on both ends If you have tried the suggestions outlined above and still have problems or questions please contact Quest Scientific ...

Страница 18: ... Hz to 4 kHz Controls BYPASS halts noise cancellation by routing input directly to output HOLD suspends adaptation to evolving noise characteristics CLEAR clears the noise replica Display LED indicates changing noise levels GREEN decreasing amplitude of the noise replica RED increasing amplitude of the noise replica QUEST SCIENTIFIC 1164 West 21st Street North Vancouver BC Canada V7P 2C8 Voice and...

Отзывы: