Quantum turret 6/FO Скачать руководство пользователя страница 14

14

 

 

3.

 

ERROR

 

CODES

 

When

 

errors

 

occur,

 

the

 

line

 

1

 

of

 

the

 

display

 

presents

 

an

 

error

 

code.

  

Line

 

3

 

of

 

the

 

display

 

identifies

 

the

 

error

 

and

 

line

 

4

 

of

 

the

 

display

 

presents

 

possible

 

solutions.

 

The

 

most

 

common

 

events

 

that

 

cause

 

errors

 

to

 

be

 

displayed

 

are

 

loose

 

cables

 

or

 

inadequate

 

coolant

 

flow.

  

For

 

errors

 

not

 

easily

 

solved,

 

please

 

contact

 

us

 

through

 

our

 

website,

 

www.qnw.com.

 

 

E5–

 

cell

 

out

 

of

 

range

 

warnings:

 

loose

 

cable,

 

sensor

 

failure

 

The

 

temperature

 

controller

 

is

 

not

 

receiving

 

a

 

reasonable

 

response

 

from

 

the

 

sensor

 

on

 

the

 

cuvette

 

tower.

  

Either

 

the

 

sensor

 

has

 

failed

 

or

 

a

 

cable

 

is

 

not

 

making

 

a

 

good

 

connection.

 

 

E6

 

 

cell

 

out

 

of

 

range

 

warnings:

 

loose

 

cable,

 

check

 

cable

 

The

 

temperature

 

controller

 

is

 

not

 

receiving

 

reasonable

 

responses

 

from

 

either

 

the

 

cell

 

tower

 

or

 

heat

 

exchanger

 

sensors.

  

Since

 

it

 

is

 

very

 

unlikely

 

for

 

both

 

to

 

fail,

 

probably

 

a

 

cable

 

is

 

loose.

 

 

E7

 

 

heat

 

exchanger

 

error

 

warnings:

 

loose

 

cable,

 

sensor

 

failure

 

The

 

temperature

 

controller

 

is

 

not

 

receiving

 

a

 

reasonable

 

response

 

from

 

the

 

sensor

 

on

 

the

 

heat

 

exchanger.

  

Either

 

the

 

sensor

 

has

 

failed

 

or

 

a

 

cable

 

is

 

not

 

making

 

a

 

good

 

connection.

 

 

E8

 

 

inadequate

 

coolant

 

warnings:

 

inadequate

 

coolant,

 

water

 

temperature

 

The

 

sensor

 

on

 

the

 

heat

 

exchanger

 

is

 

reading

 

a

 

temperature

 

above

 

60

 

°C.

  

Temperature

 

control

 

has

 

been

 

shut

 

down

 

to

 

prevent

 

damage

 

to

 

the

 

Peltier

 

element.

  

Either

 

the

 

water

 

was

 

too

 

warm

 

or

 

the

 

rate

 

of

 

flow

 

was

 

inadequate

 

to

 

draw

 

sufficient

 

heat

 

from

 

the

 

heat

 

exchanger.

 

 

 

Содержание turret 6/FO

Страница 1: ...99019 8606 Tel 509 624 9290 Fax 509 624 9488 E mail quantum qnw com Web www qnw com turret 6 FO 6 Position Temperature Controlled Cuvette Turret for Fiber Optic Spectroscopy Manual Product Overview t...

Страница 2: ...the turret 6 10 A General Description 10 B Cuvettes 11 C Cuvette z Height 11 D Cooling Water for the Peltier 11 E Dry Gas Purging 12 F External Temperature Probe 12 G External Computer Control 12 H De...

Страница 3: ...le 5 USB cable 6 turret 6 accessory kit slits wrench blanks and 6 magnetic stir bars 7 vinyl tubing to connect water and gas Depending on Your Order Your Shipment May Contain Lens Systems 1 two t6 QCL...

Страница 4: ...ure that water is flowing back into the bucket Check for leaks 4 If you plan to work at low temperatures connect a source of dry gas typically nitrogen to release on the cuvette windows Attach a lengt...

Страница 5: ...osition 2 The collimating optics produce a 5 mm diameter beam that passes through the cuvette centered at 8 5 mm above the outside bottom surface Attach fiber optics as describe below Typically light...

Страница 6: ...3 to reflect the excitation source back on itself Figure 3 Components of a lens assembly 1 Use the 5 64 inch hex screwdriver from the turret 6 accessory kit to loosen the set screw Figure 3 2 Remove t...

Страница 7: ...urret 6 insert the CD into the drive If the installation does not start automatically locate the SETUP exe file in the root directory and run it The installation process starts with a small black wind...

Страница 8: ...Start page is being displayed press SET to turn off temperature control 7 After measurements are completed turn off power on the back of the TC 1 controller and turn off the water source F Using the M...

Страница 9: ...and press SET With ramping set turning on temperature control will generate a linear ramp to the target temperature When the sample holder reaches the target temperature the ramp rate will be set back...

Страница 10: ...tte to drive rotating magnet Dry gas enters a 1 16 inch hose barb on the front of the instrument and is carried through a manifold in the floor of the turret to small holes on the bottom of the 18 cuv...

Страница 11: ...ng it is particularly important to transfer this heat away from the Peltier This is accomplished with flowing water through the heat exchanger A source of water or other cooling fluid must be connecte...

Страница 12: ...pt the plug on a standard Series 400 or Series 500 thermistor probe When a probe is plugged into the jack the probe temperature is presented on the display of the TC 1 Place the probe in a sample to m...

Страница 13: ...he source fiber into the middle of the cuvette with a magnification of 2 7 The emission lens t6 QIL em focuses a small region of the illuminated volume in the center of the cuvette onto the end of the...

Страница 14: ...out of range warnings loose cable check cable The temperature controller is not receiving reasonable responses from either the cell tower or heat exchanger sensors Since it is very unlikely for both...

Страница 15: ...ature range 20 to 110 C temperature precision 0 02 C cuvette size outside dimensions 12 5 x 12 5 mm minimum cuvette height 30 mm cuvette z height 15 mm optical port dimensions 12 mm high x 10 mm wide...

Страница 16: ...rom the set temperature of temperature readings returned from the temperature controlled turret 6 for set temperatures between 15 C and 110 C Thus it is a measure of the ability of the cuvette holder...

Страница 17: ...reproducibility is a measure of the ability of the temperature to return to an original value See Figure 1 It accounts for differences depending on the direction of temperature change and differences...

Страница 18: ...hown the time required to get within 1 C the time to get within the lock on temperature of 0 05 C from the target and the final time indicating that the TC 1 indicates the temperature is stable Temper...

Страница 19: ...e controller In this case the controller includes electronics which convert the USB connection to a serial communications port However for the port to be available to programs on the computer it will...

Страница 20: ...and set it to the most recent non zero stirrer speed setting F1 SS Turn stirrer off F1 SS What is the current stirrer speed setting F1 SS 1000 Stirrer speed setting is 1000 rpm 4 Temperature Control...

Страница 21: ...22 84 The current temperature is 22 84 C F1 CT 3 Periodically report current temperature every 3 seconds F1 CT Stop periodic current temperature reports 8 Error Reporting F1 ER Report the current erro...

Страница 22: ...re is 30 5 degrees F1 PA Stop automatic reporting of probe temperature every temperature increment F1 PX Change probe temperature returned to a precision of 0 01 degree 10 Temperature Ramping F1 RR S...

Страница 23: ...anger Temperature F1 HL What is the high temperature limit for the heat exchanger F1 HT 60 The heat exchanger high temperature limit is 60 C F1 HT What is the current temperature of the heat exchanger...

Страница 24: ...een made The commands in this section are intended to allow the program to prevent the TC 1 from sending those automatic reports F1 XX R Stop R or start R reporting all manual changes including those...

Отзывы: