![Quantum Northwest TC 1 Скачать руководство пользователя страница 38](http://html.mh-extra.com/html/quantum-northwest/tc-1/tc-1_user-manual_770893038.webp)
9/17/2020
Page 38 of 44
QNW TC 1 User’s Guide
6.14. Temperature Control Status
When temperature control is off, the circle will be a steady gray and the label will read "off".
When temperature control is on but the sample holder temperature is not stable, the small circle will alternate
slowly between green and gray and the label to it's right will read "seeking".
When temperature control is on and the sample holder temperature is stable, the circle will be a steady green
and the label will read "holding".
If an error condition occurs that causes temperature control to be turned off the circle and the caption to its
right will alternate rapidly between red and gray.
6.15. Heat Exchanger Temperature
The temperature of the heat exchanger in the sample holder is displayed here.
The heat exchanger temperature is dependent on the temperature and the rate of flow of the water being
circulated through the sample holder and on the temperature control settings.
As a warning, when the heat exchanger temperature is within 10 °C of the preset limit (typically 60 °C, the
color of the label will change to orange. To correct the problem, you must reduce the temperature of the
water (for example, by adding ice or cold water to the coolant container).
If the heat exchanger temperature exceeds that preset limit, temperature control will automatically be turned
off to prevent damage to the system.
If temperature control does turn off, the label will turn red and an error message will display below the menu
bar as illustrated below; also, the
Temperature Control Status
will show the error number and begin flashing
red. The error message (and the flashing of the status control) can be removed at any time by clicking on
the Clear button to the left of the message.
Once the problem with the heat exchanger temperature is corrected, you can restart temperature control by
clicking on the
Temperature Control On/Off checkbox
or using the menu item
Tools>(Temperature
Control)>On
.