QNW TC 1 User’s Guide
For many of the commands listed below, the controller returns information in response to the command. All
commands and responses are delineated by left and right square brackets ( [ ] ). Any text sent to the
controller not enclosed within brackets will be ignored. In this document, an ellipsis (…) is used to distinguish
responses from commands.
Note:
The format of the commands is shown below.
[command]
Purpose of the command (sent to the controller).
...[response]
Meaning of the response (received from the controller).
5.1.
Sample Holder ID Number
[F1 ID ?]
What is the ID number of the sample holder being controlled?
...[F1 ID 14]
ID is 14.
Assigned Identities:
ID = 00
specialty sample holder
(see command class 14)
14
single
24
dual
34
turret or linear multi-sample holder
5.2.
Controller Firmware Version
[F1 VN ?]
What is the
V
ersion
N
umber of the controller firmware?
...[F1 VN 2.22]
The controller firmware version number is 2.22.
5.3.
Stirrer
[F1 MS ?]
What is the
M
aximum stirrer
S
peed?
...[F1 MS 2500]
The maximum stirrer speed permitted is 2500 rpm.
[F1 LS ?]
What is the
L
owest stirrer
S
peed?
...[F1 MS 300]
The lowest stirrer speed permitted is 300 rpm.
[F1 SS S 1000]
Set
S
tirrer
S
peed to 1000 rpm and turn stirring on.
[F1 SS S 0]
Turn stirrer off (does not change the speed setting).
[F1 SS +]
Turn stirrer on and set it to the most recent non-zero stirrer speed setting.
[F1 SS -]
Turn stirrer off (this does not change the speed setting)..
[F1 SS ?]
What is the current stirrer speed setting? Depending on the R+/R- state as
specified below, the stirrer status (
’+’
/on or
’-’
/off) may also be reported.
...[F1 SS 1000]
Stirrer speed setting is 1000 rpm.