![Quantum Northwest TC 1 Скачать руководство пользователя страница 19](http://html.mh-extra.com/html/quantum-northwest/tc-1/tc-1_user-manual_770893019.webp)
9/17/2020
Page 19 of 44
QNW TC 1 User’s Guide
3.4.
Troubleshooting
If the submersible pump has operated without being immersed in water, it may not prime properly. If this
occurs, place one drop of detergent on the pump intake and try again.
If the
TC 1
display indicates that there is inadequate coolant, it may be that the circulating water is too warm.
Add ice to the water to cool it down.
If a circulating bath is used as the water source, it may not be possible to maintain a flow rate of 200 mL per
minute. While this condition will limit the lowest temperature attainable, it may be partially compensated by
using colder circulating liquid. Below about -25 °C, 30% methanol in water is generally recommended as the
coolant.
If the set temperature is near room temperature and too close to the temperature of the cooling water, the
sample holder temperature may oscillate around the set temperature. In this case, changing the temperature
of the cooling water (for example by adding ice) should stop the oscillation. Cooler circulating water can
result in improved performance of the cuvette holder at low temperatures, whereas warmer water can
improve performance at very high temperatures.
If a cuvette shorter than 44 mm is used, it may be difficult to remove. Use a taller cuvette instead.
3.5.
Error Messages
When errors occur, the line 1 of the
TC 1
Temperature Controller display presents an error code. Line 3 of
the display identifies the error and line 4 of the display presents possible solutions. The most common events
that cause errors to be displayed are loose cables or inadequate coolant flow. For errors not easily solved,
contact QNW at www.qnw.com/contact-us/ or [email protected].
Specific error messages include the following.
5 – cell out of range
warnings:
loose cable, sensor failure
The temperature controller is not receiving a reasonable response from the sensor on the cuvette tower.
Either the sensor has failed or a cable is not making a good connection.
6 – cell out of range
warnings:
loose cable, check cable
The temperature controller is not receiving reasonable responses from either the cell tower or heat exchanger
sensors. Since it is very unlikely for both to fail, probably a cable is loose.
7 – heat exchanger error
warnings:
loose cable, sensor failure
The temperature controller is not receiving a reasonable response from the sensor on the heat exchanger.
Either the sensor has failed or a cable is not making a good connection.