19
Rev 04-2009
SERVICE MANUAL
If no biometry and/or ProBeam parameters have been installed, you
have to:
Click on the
« Add »
icon,
Select the CD-Rom drive,
Select the
« .prb »
file,
Click on
« Open »
, the probe is added in the list.
If the biometry and/or ProBeam parameters are installed, you have just
to perform the calibration of each probe:
Select the probe to calibrate in the list, (the selected probe is
highlighted in blue),
Click on the
« Test Block »
key to perform the probe calibration of the
selected probe and get the correct position of the first marker.
5- MAINTENANCE
The following window is opened, by clicking on the « Test Block » key: the Zero
(Probe) value is set at 0.16mm by default.
Содержание Aviso