SERVICE MANUAL
5- MAINTENANCE
5-1 TOUCH SCREEN CALIBRATION
The position of the touch-screen-sensitive area can be calibrated.
It can be necessary after a transport, a screen assembly changing, a
software upgrade or other wich could change the touch screen calibration.
This calibration can be done only if the Aviso software on thce computer is
not started or if the Aviso and the computer are not linked.
Switch ON the Aviso and wait for the home page to display.
Turn the power knob clockwise until '
Calibration
' displays on the screen.
NOTE :
Read the entire procedure before to start.
1-
Press the ‘
START
’
button.
2-
Wait for the first beep;
it is the begining of the calibration
procedure.
3-
Press ‘
P
' until to hear the
second beep.
4-
Then press successively the 4
screen corners.
Each time a corner is pressed,
repetitive beeps will be hear.
It means that the screen pressure is
correctly recognized.
Go to the following corner.
NOTE :
To obtain a better result, it is recommended to press the screen
with a thin plastic stick, like a PDA pencil.
5-
Press '
P
' again to complete the procedure.
6-
At the end, you can save the new calibration and exit by ‘
SAVE
’ (the
screen comes back to the home page), abort the procedure by ‘
ESC
’
or start it again by ‘
START
’.