QuantAsylum QA451 User’s Manual

 

 

 

 

Page 9 

 

 

 

QA451 Software 

The QA451 and QA451B 

will appear on your PC as a “HID” device with a USB VID:PID of 0x16C0:0x4E

35. 

It doesn’t need drivers. It will show up in the Device Manager as a “HID

-compliant vendor-defined 

device.”

 

When you first plug in the QA451 device, Windows should spend a bit of time configuring the device. 
Download and install the QA451 application fro

Github

. After launching the application, you should 

see the following UI: 

 

In the title bar, you can see the version of the software, shown here as 1.0. In the status bar, we can see 
the hardware 

device is “opened” 

and that this is firmware version 30. You will be alerted if the firmware 

version is incorrect and given a chance to re-flash the QA451 hardware. It is important to NOT use the 
application if the firmware version is reported as incorrect as malfunctions could occur.  

When the QA451 

and application are “talking”, should see the “LINK” LED on the 

QA451 hardware 

illuminated.  

If 

you press the “8 Ohms” button, you will hear a “click” from the 

QA451 relays and the application will 

report the 8 ohm load is active as shown below. At this point, you can use a DVM to measure across the 
L+ and L- Load Input pins and confirm that 8 ohms is being presented. You can repeat the experiment for 
4 ohms, too, and also for the right channel. 

 

Содержание QA451

Страница 1: ...QuantAsylum QA451 User s Manual Page 1 QuantAsylum QA451 B User s Manual Programmable 4 8 Ohm Load SW Release 1 4 October 2023...

Страница 2: ...and the voltage and current limits for each input See additional safety notices throughout this document Limited Warranty This product has a limited warranty for 6 months from the time of purchase Dur...

Страница 3: ...uts 7 DUT Power Soft Start 7 Load Inputs 8 BNC Outputs 8 QA451 Software 9 Application Temperature Sensing 10 Application Current Sensing 10 DUT Power 10 Fault Conditions 10 Remote Connections 10 QA451...

Страница 4: ...QuantAsylum QA451 User s Manual Page 4 Tractor 19...

Страница 5: ...are electrically isolated from the USB grounds This isolation is limited to 50V Do not connect the QA451 to a product that has its ground reference more than 50V from the PC ground Current Sense Input...

Страница 6: ...high side PMOS switch Power handling up to 300W duration limited Current limiting of supply for accidental shorts trips when 15A for more than 2mS Soft start allowing gentle turn on of the amplifier D...

Страница 7: ...o be used For example if you are testing Class D amplifier boards it might make sense to use the current sense and power the boards from a fixed external supply But if you are testing finished amplifi...

Страница 8: ...not exceed the 200W for 220 mS rating of the Amplifier Load subsystem BNC Outputs The BNC outputs are single ended attenuated version of the load inputs The attenuation is 12 dB This means if your amp...

Страница 9: ...us bar we can see the hardware device is opened and that this is firmware version 30 You will be alerted if the firmware version is incorrect and given a chance to re flash the QA451 hardware It is im...

Страница 10: ...system The resolution is around 5 to 10 mA That is you will see readings jump by perhaps 5 mA On the QA451B the resolution is around 10 to 20 mA For high power amplifiers that might draw 10A or so thi...

Страница 11: ...s port 9450 opened then run the QA451 application with admin rights and see if that allows a browser to open the link above If that solves the problem then you can either run the app with admin rights...

Страница 12: ...to handle 25W when the case temperature is 25C That 25W figure is de rated by about 15 as the backplate temperature reaches 50C If the resistors are near room temperature and the QA451 is configured f...

Страница 13: ...ility of the QA401 and QA451 combine to allow thorough testing at high power levels The QA451 is obviously not the platform you d chose for burning in amplifiers But a test suite that has an average p...

Страница 14: ...on the switch contact us at the support alias and we can work to help solve the problem Thermal Considerations The QA451 is NOT designed for sinking 10 s of watts for long periods of time The QA451 c...

Страница 15: ...ge power drops to 1 5W In summary you need to plan your thermal budget ahead of time to ensure the QA451 can meet your needs Hitting the thermal targets above will ensure continuous operation without...

Страница 16: ...QuantAsylum QA451 User s Manual Page 16 Appendix I Typical Connection Diagram...

Страница 17: ...itch is on S will be equal to S The switch is not bidirectional If S exceeds S then current will begin to flow through the MOSFET body diode 2 Soft start is provided to allow power supplies with large...

Страница 18: ...QuantAsylum QA451 User s Manual Page 18...

Страница 19: ...nput Gain option on the QA401 This is how you tell the QA401 application that the value it is measuring is less than the actual value and thus the reported peaks and noise for example must be increase...

Страница 20: ...QuantAsylum QA451 User s Manual Page 20...

Отзывы: