QuantAsylum QA451 Скачать руководство пользователя страница 7

QuantAsylum QA451 User’s Manual

 

 

 

 

Page 7 

 

 

 

8 Ohm LED

  

 

This LED indicates the QA451 is presenting an 8 ohm load at the load inputs 

4 Ohm LED

  

 

This LED indicates the QA451 is presenting a 4  ohm load at the load inputs 

DUT Power LED

 

This LED indicates if the high-side switch is enabled or not. If the LED is flashing 
rapidly, then soft-start is in process. 

If both the 8 and 4 ohm LEDs are flashing, then it means the QA451 has reached a thermal limit. This is 
not harmful to the QA451. When this happens, the QA451 will open the relays automatically and ensure 
the load inputs are removed from the internal load resistors. 

Current Sense Inputs 

The QA451 uses dual-range isolated Hall sensing device to measure the power supply DC. The positive 
terminal of the amplifier

’s

 DC power supply flows into the S+ pin on the front panel, is sensed by the Hall 

sensing devices, and then flows out of the S- terminal to the amplifier (see Appendix I Connection 
Diagram). The Hall sensors used on the QA451 is an Allegro Micro ACS711, with +/- 25A of measurement 
range, and an ACS70331 with +/-2.5A of measurement range. On the QA451B, a single Ti 
TMSCS1108A1U) is used. On the QA451, only positive current (from DC supply to the amp) are reported. 
If a negative current is measured, it is clamped at 0 amps. 

The DC switch on the QA451 is designed to operate continuously at 50V with 10A of load and can 
withstand normal transients excursions above 50V. 

The current sensing is not required to be used. For example, if you are testing Class D amplifier boards, it 
might make sense to use the current sense and power the boards from a fixed external supply. But if 
you are testing finished amplifiers that runs from AC, then the DC measurement provided by the QA451 

isn’t useful and can be skipped

ATTENTION! DO NOT CONNECT THE CURRENT SENSE CONNECTOR WHEN THE DUT POWER LED IS ON! 

Do not swap amplifier boards when the DUT POWER LED is on! 

If your power supply and amplifier both have massive capacitors, and you connect the amplifier while 
the DUT power is enabled, then potentially hundreds or even thousands of amps can flow through the 
high-side switch and cause it to fail as the charge is re-distributed  

The Current Sense connector on the QA451 is a Molex 39536-0002. The mating connector is a Molex 
39534-0002. This connector has a rating of 300V and 15A per pin.  

Do not exceed the 10A RMS rating or 15A max rating of the Current Sense subsystem. 

DUT Power Soft Start 

When the DUT Power is activated, the DUT Power LED will flash rapidly for 1 second. This indicates that 
soft-start is underway. During soft-start the DUT amplifier is connected to the high-side rail (S+) through 
a 10 ohm resistor. This allows amplifier

s with very large filter caps to charge at current levels that won’t 

stress components in the power supply, the QA451 and the DUT amplifier. A large amplifier might have 
10 mF of bulk capacitance. The RC time constant of the 10 mF and 10 ohm is 100 mS. Thus, the one-
second soft-start allows 10 time constants of soft-start to elapse before bypassing the soft-start resistor. 
See the discussion below on 

Faults

 for more analysis.  

Содержание QA451

Страница 1: ...QuantAsylum QA451 User s Manual Page 1 QuantAsylum QA451 B User s Manual Programmable 4 8 Ohm Load SW Release 1 4 October 2023...

Страница 2: ...and the voltage and current limits for each input See additional safety notices throughout this document Limited Warranty This product has a limited warranty for 6 months from the time of purchase Dur...

Страница 3: ...uts 7 DUT Power Soft Start 7 Load Inputs 8 BNC Outputs 8 QA451 Software 9 Application Temperature Sensing 10 Application Current Sensing 10 DUT Power 10 Fault Conditions 10 Remote Connections 10 QA451...

Страница 4: ...QuantAsylum QA451 User s Manual Page 4 Tractor 19...

Страница 5: ...are electrically isolated from the USB grounds This isolation is limited to 50V Do not connect the QA451 to a product that has its ground reference more than 50V from the PC ground Current Sense Input...

Страница 6: ...high side PMOS switch Power handling up to 300W duration limited Current limiting of supply for accidental shorts trips when 15A for more than 2mS Soft start allowing gentle turn on of the amplifier D...

Страница 7: ...o be used For example if you are testing Class D amplifier boards it might make sense to use the current sense and power the boards from a fixed external supply But if you are testing finished amplifi...

Страница 8: ...not exceed the 200W for 220 mS rating of the Amplifier Load subsystem BNC Outputs The BNC outputs are single ended attenuated version of the load inputs The attenuation is 12 dB This means if your amp...

Страница 9: ...us bar we can see the hardware device is opened and that this is firmware version 30 You will be alerted if the firmware version is incorrect and given a chance to re flash the QA451 hardware It is im...

Страница 10: ...system The resolution is around 5 to 10 mA That is you will see readings jump by perhaps 5 mA On the QA451B the resolution is around 10 to 20 mA For high power amplifiers that might draw 10A or so thi...

Страница 11: ...s port 9450 opened then run the QA451 application with admin rights and see if that allows a browser to open the link above If that solves the problem then you can either run the app with admin rights...

Страница 12: ...to handle 25W when the case temperature is 25C That 25W figure is de rated by about 15 as the backplate temperature reaches 50C If the resistors are near room temperature and the QA451 is configured f...

Страница 13: ...ility of the QA401 and QA451 combine to allow thorough testing at high power levels The QA451 is obviously not the platform you d chose for burning in amplifiers But a test suite that has an average p...

Страница 14: ...on the switch contact us at the support alias and we can work to help solve the problem Thermal Considerations The QA451 is NOT designed for sinking 10 s of watts for long periods of time The QA451 c...

Страница 15: ...ge power drops to 1 5W In summary you need to plan your thermal budget ahead of time to ensure the QA451 can meet your needs Hitting the thermal targets above will ensure continuous operation without...

Страница 16: ...QuantAsylum QA451 User s Manual Page 16 Appendix I Typical Connection Diagram...

Страница 17: ...itch is on S will be equal to S The switch is not bidirectional If S exceeds S then current will begin to flow through the MOSFET body diode 2 Soft start is provided to allow power supplies with large...

Страница 18: ...QuantAsylum QA451 User s Manual Page 18...

Страница 19: ...nput Gain option on the QA401 This is how you tell the QA401 application that the value it is measuring is less than the actual value and thus the reported peaks and noise for example must be increase...

Страница 20: ...QuantAsylum QA451 User s Manual Page 20...

Отзывы: