One thing to keep in mind is that the Measurement Stop value sets the upper harmonic to be
considered. If you are looking at a 1 KHz signal, then a 20 KHz stop value means you will look at all
harmonics up to the 20
th
harmonic in your THD calculation.
But what if you have specified an 11 KHz signal? In that case, the second harmonic will be at 22 KHz,
which is beyond the 20 KHz Measurement Stop limit. In that case, the second harmonic will never be
seen, and the THD value won’t make much sense. In the example below, we can see a marker at 11 KHz
and 22 KHz, and since the Measurement Stop is set to 20 KHz the second harmonic is never seen.
THD Context Menu
The THD context menu is very simple, and allows you to set the Measurement Start and Stop limits.
Note these are the same limits that are set in the Power context menu. Additionally, you can specify
what will be used for the fundamental.