For Research Use Only
Chapter 1: Installation
and Safety Information
SAFETY PRECAUTIONS
Lid and Glass/Tray Insert
Make sure the lid is open and is not moving when
placing the plate in the tray. Also, if removing the
glass/tray insert for cleaning, exercise caution that
nothing is dropped on the lens of the camera.
Electricity
S
ince the Q-View Imager LS has a 110V/220V power
supply, exercise typical precautions to ensure shocks
do not occur. Do not place liquids close to the Q-View
Imager LS and do not cut cords that provide power or
connect the device to the PC.
PLACEMENT IN THE
LABORATORY
Temperature and Humidity
The Q-View Imager LS has a high resolution camera,
which may be subject to performance failures if the
temperature is too high. If the lab temperature is too
high, the camera may be unable to perform correctly.
Recommended operating conditions are 15–35°
C. Humidity can also cause operation problems
for the Q-View Imager LS. Ensure your laboratory
environment does not exceed a dew point of 20° C to
prevent condensation.
Ventilation and Dust
Ensure the room where the Q-View Imager LS is
placed has adequate ventilation to ensure passive
cooling can occur. Dust can collect on the glass on