Quansys Q-View IMAGER LS Скачать руководство пользователя страница 14

10

For Research Use Only

Chapter 3: Operation

BEFORE YOU BEGIN

The following procedure is recommended before 

every use.

1. 

Clean the top of the glass/tray insert and ensure 

it is free of debris.  

2.  Check within Q-View Software that the Q-View 

Imager LS is connected to the PC. 

CLEANING THE TRAY SURFACE

If  any  particulate,  liquid,  or  dried  salt  residues  or 

smudges  are  left  on  the  glass  or  the  tray,  it  will 

adversely  affect  the  quality  of  the  image  and  the 

overall quality of your data. This can be cleaned with 

70 percent ethanol using a lint-free cloth. Do not use 

scouring cleaners or abrasive pads as this can scratch 

the glass.

LOADING THE PLATE

The user sets the Q-Plex™ plate in the glass/tray 

insert with well A1 in the top left corner of the tray 

when facing the Q-View Imager LS. Make sure the 

plate rests flat on the glass on all edges of the plate. 

Make sure there is no lid on the plate and there is 

only one plate on the tray. Then, gently close the lid 

and proceed to image.

Содержание Q-View IMAGER LS

Страница 1: ...Version 2 03 For Research Use Only Not For Use in Diagnostic Procedures Operator s Manual...

Страница 2: ...h the device instruction manual It is critical that the user read this manual in its entirety and train on the device frequently to ensure proper operation For laboratory use only Pollution Degree 2 o...

Страница 3: ...ANSYS BIOSCIENCES BRANDED PRODUCTS ANY WARRANTY APPLICABLE TO NON QUANSYS BIOSCIENCES BRANDED SOFTWARE OR PRODUCTS IS PROVIDED BY THE ORIGINAL MANUFACTURER Printing History Version Number Date 1 0 Sep...

Страница 4: ...nstallation and Safety Information 2 Safety Precautions 2 Lid and Glass Tray Insert 2 Electricity 2 Placement in the Laboratory 2 Temperature and Humidity 2 Ventilation and Dust 2 Space 3 Vibration 3...

Страница 5: ...al Features 8 Back 8 Computer Connection 8 Q View Software Introduction 8 Chapter 3 Operation 9 Before you Begin 9 Cleaning the Tray Surface 9 Loading the Plate 9 Imaging the Plate 9 Using the Qualifi...

Страница 6: ...ot cut cords that provide power or connect the device to the PC PLACEMENT IN THE LABORATORY Temperature and Humidity The Q View Imager LS has a high resolution camera which may be subject to performan...

Страница 7: ...t cause vibrations such as vortex mixers shakers centrifuges etc are not co located on the bench with the Q View Imager LS Also consider where the Q View Imager LS is placed in the laboratory For exam...

Страница 8: ...and a dark field acquired Focusing the Q View Imager LS Open the Q View Imager LS lid Place the 96 well focusing plate included on the plate tray From the Manage Imagers menu select Focus With the li...

Страница 9: ...onnection on any computer running Q View Software goes through a proxy specify these settings under Administration Network Settings Your Q View Software and Q View Imager LS are now ready to use Disas...

Страница 10: ...protective sleeve 9 Bundle and tie cables together 10 Insert the Q View Imager LS into the foam insert in the Q View Imager LS box 11 Place the glass tray insert and cords into the box CHEMICAL SAFET...

Страница 11: ...reason please contact technical support Do not use an extension cord for the Q View Imager LS Ensure that all equipment plugged into the same circuit as the Q View Imager LS does not exceed the ampere...

Страница 12: ...ray insert is free from dust and any chemicals To clean the device use 70 percent ethanol or warm water Do not use scouring cleaners or abrasive pads as this can scratch the glass Be careful not to sp...

Страница 13: ...Management and see an entry in a table for the Q View Imager LS with one of the columns saying connected or disconnected Q VIEW SOFTWARE INTRODUCTION A copy of Q View Software will accompany the Q Vie...

Страница 14: ...tray it will adversely affect the quality of the image and the overall quality of your data This can be cleaned with 70 percent ethanol using a lint free cloth Do not use scouring cleaners or abrasive...

Страница 15: ...can be used to qualify the performance of the Q View Imager LS over time This plate has built in LEDs that create a constant signal that can be imaged over certain periods of time with Q View Imagers...

Страница 16: ...Solution Q View Software is unable to recognize the Q View Imager LS Failure of communication between the Q View Imager LS and computer Ensure power is supplied to the Q View Imager LS Ensure the USB...

Страница 17: ...nescent blots 7 6 cm x 12 1 cm Imaging Time 1 5 minutes Control Q View Software Connection to PC USB OPERATING SPECIFICATIONS Operating Conditions Temperature range of 15 35 C and dew point of no grea...

Страница 18: ...Part Number 365 North 600 West Logan Utah 84321 Tel 1 888 782 6797 Fax 435 750 6869 www quansysbio com For technical support email support quansysbio com...