Page 82 of 85
Service & Calibration
4.3.6
AC Current Calibration (ACA)
Connect the OUTPUT terminal of the Sentry Plus Series instrument to a resistance box or
resistance standard. Connect an AC/DC current meter in series between the resistance load
(box/standard) and the RTN/LOW terminal of the Sentry Plus instrument. Table 4-3 lists the
resistance loads necessary for the current calibration steps.
Table 4-3: Resistance Loads
Mode Step
#
Voltage
Calibration
Point
Resistance
(Load)
ACA
CAL 7
1200 V
0.12mA
10M
Ω
ACA
CAL 8
1200 V
2.4mA full
500k
Ω
ACA
CAL9
1200 V
2.4mA
500k
Ω
ACA
CAL 10
1200 V
12mA
100k
Ω
DCA
CAL 11
1200 V
0.12mA
10M
Ω
DCA
CAL 12
1200 V
2.4mA full
500k
Ω
DCA
CAL 13
1200 V
2.4mA
500k
Ω
DCA
CAL 14
1200 V
4.8mA
250k
Ω
Press [F1] = UP to go to
Cal Step 7: ‘ACA 3mA Offset (0.12mA)’
.
Press [STOP] to return instrument to steady state.
Press [START] to get offset value.
Press [F1] = INC or [F2] = DEC until the SPlus display reads the same as the current meter.
Press [ENTER] to accept reading.
Sentry Plus reverts to “ACA 3mA Offset” display (incorporating the new cal value).
Press [F1] = UP to go to
Cal Step 8: ‘ACA 3mA Full (2.4mA)’
.
Press [STOP] to return instrument to steady state.
Press [START] to get full value.
Press [F1] = INC or [F2] = DEC until the SPlus display reads the same as the current meter.
Press [ENTER] to accept reading.
Sentry Plus reverts to “ACA 3mA Full” display (incorporating the new cal value).
Press [F1] = UP to go to
Cal Step 9: ‘ACA 15mA Offset (2.4mA)’
.
Press [STOP] to return instrument to steady state.
Press [START] to get offset value.
Press [F1] = INC or [F2] = DEC until the SPlus display reads the same as the current meter.
Press [ENTER] to accept reading.
Sentry Plus reverts to “ACA 30mA Offset” display (incorporating the new cal value).
Press [F1] = UP to go to
Cal Step 10: ‘ACA 15mA Full (12mA)’
.
Press [STOP] to return instrument to steady state.
Press [START] to get full value.
Press [F1] = INC or [F2] = DEC until the SPlus display reads the same as the current meter.
Press [ENTER] to accept reading.
Sentry Plus reverts to “ACA 30mA Full” display (incorporating the new cal value).