Operation
Page 39 of 85
2.6
Programming a DC Hipot Test
This test is applicable to the Sentry 20 and 30 Plus instruments. With the instrument in ‘stand-
by’ status, press [F1] = PROGRAM. If continuing the 5-step example from ¶2.5, you are already
in program mode and on the AC test mode page. The example illustrated herein shows a GC test
as Step 1, an AC test as Step 2 and how to program a DC test in Step 3. Follow the
green
arrows
(
!
) on the right side of this diagram to program the individual DC hipot test parameters.
Continue on next page.
INC.
DEC.
EXIT
NEXT
RMT
ERR
OFST
LOCK
0.05 - 6kV
F1
F4
F3
F2
FALL
RAMP
ARC
LOW
DC
STEP 3/3
VOLT
HIGH
TIME
Select
DC
Test Mode
Set
Test Voltage
Set
High Current
Limit
To set high current limit:
0.0001 - 5mA in .001mA increments
To move highlighted box to HIGH.
To move highlighted box to TIME.
INC.
DEC.
EXIT
NEXT
RMT
ERR
OFST
LOCK
0.0001 - 5mA
F1
F4
F3
F2
FALL
RAMP
ARC
LOW
DC
STEP 3/3
VOLT
HIGH
TIME
INC.
DEC.
EXIT
ENTER
RMT
ERR
OFST
LOCK
0, 0.1 - 999s
F1
F4
F3
F2
FALL
RAMP
ARC
LOW
DC
STEP 3/3
VOLT
HIGH
TIME
INC.
DEC.
EXIT
NEXT
RMT
ERR
OFST
LOCK
0 - 5mA 0 = OFF
F1
F4
F3
F2
FALL
RAMP
ARC
LOW
DC
STEP 3/3
VOLT
HIGH
TEST
To set test time:
0, 0.1 - 999s in 0.1s increments
To move highlighted box to LOW.
Set
Test Time
To move highlighted box to ARC.
To set low current limit:
0 - 5mA in 0.001mA increments
To set test voltage:
0.05 - 6kV in .001kV increments
OFF
OFF
OFF
OFF
OFF
OFF
OFF
OFF
OFF
OFF
OFF
OFF
OFF
OFF
OFF
OFF
To move highlighted box to VOLT
Set MODE = DC
UP
DOWN
EXIT
NEXT
RMT
ERR
OFST
LOCK
Select Mode
F1
F4
F3
F2
FALL
RAMP
ARC
LOW
AC
STEP 3/3
VOLT
HIGH
TIME
OFF
OFF
OFF
OFF
[F1] = New
Program
Step 3 = DC
To move highlighted box around
display to select parameter to change
To exit programming mode.
To delete or insert a test step.
To change value in highlighted box.
FIRST
MORE..
EXIT
NEXT
RMT
ERR
OFST
LOCK
Voltage is 0
F1
F4
F3
F2
FALL
RAMP
ARC
LOW
AC
STEP 3/3
VOLT
HIGH
TIME
OFF
OFF
OFF
OFF
0.000kV
:
:
:
:
:
:
:
0.500mA
3.0s
:
:
:
:
:
:
:
3.0s
0.500mA
0.000kV
0.000kV
:
:
:
0.500mA
3.0s
:
:
:
:
:
:
:
:
:
:
:
2.500kV
0.500mA
3.0s
:
:
:
3.0s
2.999mA
2.500kV
:
:
:
:
Set
Low Current
Limit
2.500kV
:
:
:
5.0s
2.999mA
:
:
:
:
OFF
:
I-RUS
I-RUS
I-RUS
I-RUS
:
:
:
:
OFF
DWELL
DWELL
:
OFF
OFF
OFF
OFF
DWELL
DWELL
:
OFF
OFF
Содержание Sentry 10 Plus
Страница 2: ...Page 2 of 85...
Страница 6: ...Page 6 of 85...
Страница 10: ...Page 10 of 85...
Страница 12: ...Page 12 of 85...
Страница 14: ...Page 14 of 85...
Страница 26: ......
Страница 66: ......
Страница 76: ......