Pulsed Operation

 

Press the Remote operation switch on the front panel  
Pulse as desired using transducer pulse button switch with appropriate intensity power 
setting 
 

Footswitch Operation

 

x

  Switch front panel switch to Continuous operation 

x

  Insert footswitch plug into jack on rear of the unit 

x

  Set desired power intensity level 

x

  Pulse as desired using footswitch  

  
The sides and end of the tip must 

never

 be allowed to come in direct contact with the 

container or hard surface.  The stress resulting at the point of contact could cause fracture of 
the MICROPROBE or of the glass container.   
 
Immerse the tip of MICROPROBE at least 1 to 1½ times the tip diameter into the solution 
(without touching the bottom).   
 
Lower the tip into the solution to avoid aerosolizing and foaming.  Aerosolizing and foaming 
generally occur when the tip is not immersed far enough into the solution, or if too high of a 
power setting is used by lowering the tip in the solution, decreasing power, and reducing 
solution temperature will normally prevent aerosolizing and foaming. In severe cases, use a 
narrow test tube and plastic film over the vessel.  Push the tip through the film to sonicate. 
Be sure to clean the tip before sonicating another sample, since protein released from cell 
material acts like a wetting agent and tends to promote foaming. Once foaming occurs, 
reduce power below the cavitation level before proceeding. If the foam persists, the sample 
may have to be discarded.   
 
The highest intensity energy is concentrated directly under the tip and dissipates within ¼" 
(6mm) from the tip.  Liquids and suspended solids to be processed must circulate freely in 
this zone.  Whole tissue should first be finely divided or homogenized roughly in a 
mechanical device, such as a laboratory blender or rotary homogenizer, before being 
sonicated.  Small pieces of tissue can be sonicated whole if trapped directly under the probe 
tip.   
 

Operation with optional Plug-in Timer

 

 
The timer allows the probe to be activated for a specific period of time between 1 to 15 
minutes. At   the end of the timed cycle the probe will be inactive indefinitely (the power to 
the unit remains on, however). 
 

For timed operation: 

1. 

Turn the power control knob to zero. 

2. 

Turn the power switch on. 

3. 

Allow digital reading to stabilize at 

0.0. 

4. 

Place the probe into your sample at mid-depth. 

Содержание MICROSON XL 2000

Страница 1: ...1 MICROSON TM XL 2000 ULTRASONIC LIQUID PROCESSOR 23 5 725 6 0 18 53 Church Hill Road Newtown CT 06470 USA 203 426 0101tel 877 338 9636 toll free 203 426 7026 fax info sonicator com www sonicator com...

Страница 2: ...5 I INSTALLATION A Power Requirements 6 B Inspection 7 C Placement of Equipment 7 II OPERATION A Principle of Ultrasonics 8 B Description of Major Components 8 C Description of Controls and 8 Indicat...

Страница 3: ...pment Observe the following WARNINGS High voltage is present in the generator power supply convertor and high frequency cable Do NOT attempt to remove the generator cover or convertor case Do not touc...

Страница 4: ...or and rubber boot protector at the end of the cable should be tight to the wire with no movement possible and no cracks or frayed ends visible Do not subject the cable ends to severe bending loads wh...

Страница 5: ...nsions 17cm L x 3cm Dia 6 7 L x 1 18 Dia without probes Weight 0 2 kg 0 5 lb Materials Aluminum case and Front Driver Environmental Pollution Degree 1 Temperature Limits 50q F 110q F 10q C 43q C Barom...

Страница 6: ...6...

Страница 7: ...or to set voltage for 115V 92 140RMS or 220V 200 240 VRMS service CAUTION Do not operate a unit set for 115V service on 220V line or operate a unit set for 220V service on a 115V line WARNING The elec...

Страница 8: ...terial check it carefully for small items Report any missing components to Qsonica immediately Visually inspect all external controls indicators and surfaces to detect any damage in transit If damage...

Страница 9: ...ator also known as the power source includes the operating controls and power indicator an On Off switch and separable three wire grounded line cord with integral U S plug or Europlug fuse and high fr...

Страница 10: ...placed on the probe or on the black front driver Cautions x Never touch the tip of a live probe to your hand or skin x Always allow the unit to reach room temperature before operating x Do not operate...

Страница 11: ...cases use a narrow test tube and plastic film over the vessel Push the tip through the film to sonicate Be sure to clean the tip before sonicating another sample since protein released from cell mate...

Страница 12: ...t as necessary to process all samples CAUTION If Footswitch or Timer plug is removed from jack at rear of unit then the unit will remain on indefinitely In this event turn main power switch off to dis...

Страница 13: ...tor and convertor the RMA number and a brief explanation of the problem with the unit If possible include x A Purchase Order Number x Bill To and Ship To address x The return method of shipment usuall...

Отзывы: