background image

5

 

 

SPECIFICATIONS 

 
 
 

Generator

 

 
Dimensions: 33cm W x 19cm D x 17cm H (13" x 7.5" x 6.7") 
Weight: 4.2 kg (9.2 lb.) 
Input Voltage: 115 VAC nom. (switchable to 220 VAC nom.) 50/60 hz 
Full Load Current: 1.65 amps at 115v/ 0.8 amps at 220v 
Fuse Rating and Type: 3 amp/1.5 amp Fast acting 
Voltage Tolerance: +/- 10% nominal voltage 
Output Voltage/Frequency: 950 Vrms (maximum) 22.5 kc (nom) 
 
 
 

Convertor

 

 
Dimensions: 17cm L x 3cm Dia. (6.7" L. x 1.18" Dia.) without probes 
Weight: 0.2 kg (0.5 lb.) 
Materials: Aluminum case and Front Driver 
 
 
 

Environmental 

 
Pollution Degree 1 
Temperature Limits: 50

q

 F - 110

q

 F (10

q

 C - 43

q

 C) 

Barometric Pressure: Unlimited (Note: In high vacuum areas, additional cooling provisions may 
be needed.  Contact factory) 
 
 
 

Accessories 

 
Use only accessories and probes listed in the catalog by the manufacturer as suitable for use 
with this appliance.  Do not attempt to fabricate ultrasonic tooling or accessories unless approval 
has been obtained from the manufacturer in advance. 

 

 
 
 
 
 

Содержание MICROSON XL 2000

Страница 1: ...1 MICROSON TM XL 2000 ULTRASONIC LIQUID PROCESSOR 23 5 725 6 0 18 53 Church Hill Road Newtown CT 06470 USA 203 426 0101tel 877 338 9636 toll free 203 426 7026 fax info sonicator com www sonicator com...

Страница 2: ...5 I INSTALLATION A Power Requirements 6 B Inspection 7 C Placement of Equipment 7 II OPERATION A Principle of Ultrasonics 8 B Description of Major Components 8 C Description of Controls and 8 Indicat...

Страница 3: ...pment Observe the following WARNINGS High voltage is present in the generator power supply convertor and high frequency cable Do NOT attempt to remove the generator cover or convertor case Do not touc...

Страница 4: ...or and rubber boot protector at the end of the cable should be tight to the wire with no movement possible and no cracks or frayed ends visible Do not subject the cable ends to severe bending loads wh...

Страница 5: ...nsions 17cm L x 3cm Dia 6 7 L x 1 18 Dia without probes Weight 0 2 kg 0 5 lb Materials Aluminum case and Front Driver Environmental Pollution Degree 1 Temperature Limits 50q F 110q F 10q C 43q C Barom...

Страница 6: ...6...

Страница 7: ...or to set voltage for 115V 92 140RMS or 220V 200 240 VRMS service CAUTION Do not operate a unit set for 115V service on 220V line or operate a unit set for 220V service on a 115V line WARNING The elec...

Страница 8: ...terial check it carefully for small items Report any missing components to Qsonica immediately Visually inspect all external controls indicators and surfaces to detect any damage in transit If damage...

Страница 9: ...ator also known as the power source includes the operating controls and power indicator an On Off switch and separable three wire grounded line cord with integral U S plug or Europlug fuse and high fr...

Страница 10: ...placed on the probe or on the black front driver Cautions x Never touch the tip of a live probe to your hand or skin x Always allow the unit to reach room temperature before operating x Do not operate...

Страница 11: ...cases use a narrow test tube and plastic film over the vessel Push the tip through the film to sonicate Be sure to clean the tip before sonicating another sample since protein released from cell mate...

Страница 12: ...t as necessary to process all samples CAUTION If Footswitch or Timer plug is removed from jack at rear of unit then the unit will remain on indefinitely In this event turn main power switch off to dis...

Страница 13: ...tor and convertor the RMA number and a brief explanation of the problem with the unit If possible include x A Purchase Order Number x Bill To and Ship To address x The return method of shipment usuall...

Отзывы: