General instructions
These instructions are for the users to receive information about the line pressure regulator unit
function and design
.
Read the instructions with the
line
pressure regulator in place so that you can quickly and easily find all
the controls.
The
line
pressure regulator is part of the
medical gas pipeline system
, and it is therefore
subject to the
inspections and controls specified in EN ISO 7396-1 and national standard SIS HB 370.
It may not be put into operation until a safety inspection has been carried out and a
commissioning of
medical use
has been issued.
Make sure that the instructions are easily accessible, if necessary, more copies can be ordered from
Qmt-tech. If something is unclear, you can also call us at 0480-440200.
Ensure the following
•
The medical gases must not come into contact with oil, grease. These substances become
extremely flammable on contact with certain gases and can start explosive fires.
•
Hands, tools and equipment must be clean and grease-free.
•
In order for Qmt-tech's warranty
terms
to apply, the
cleanliness requirements must be maintained.
•
Leak detectors used on gas fittings must be free of substances that can react with oxygen. It must
also not contain ammonium compounds (ammonia) as these can cause cracking in brass details.
•
Settings and repairs must be performed by technicians from Qmt-tech ab.
•
According to SIS HB 370 and EN ISO 7396-1, alarms must be installed to the
line
pressure
regulator.