Set-up & Installation 

Read Operating Instructions Thoroughly Prior to Operation  

Safety  Precautions 

Read Operating Instructions thoroughly prior to operation.  Use only a grounded outlet that is rated for your 
model's electrical requirement. Do not modify the oven or factory control settings to operate the oven above the 
stated maximum operating temperature.  Exterior surfaces on 180E-1 models may become hot to the touch when 
operating at higher set temperatures.  Conduct periodic maintenance as required.

General Operation 

*

Chamber Loading

Position unit in its ultimate operating location.  Keep a minimum of 
3" of airspace around the unit and a minimum of 6" above the unit.  
The port hole at the top of the unit will expel a small amount of 
warm air through natural convection. This port can also be used as 
an access way for external temperature measurement of a solution 
for example.

Install adjustable shelf by placing the ends of the wire shelf bracket 
into the corresponding holes located on the inner sides of the oven 
at the desired height. Push the ends of the bracket into the holes 
until the first bends in the bracket are against the wall, then rotate 
the bracket down. Place the shelf on the brackets. 

(FIG 1)

Plug the unit into a grounded outlet for your unit's rated voltage.

The unit is ready for your immediate use. All control parameters, 
calibration and tuning has been done at the factory, no adjustments 
are necessary.

Push the illuminated power button. All LED's on the temperature 
control will light up for 5 seconds until the current or actual chamber 
temperature is displayed.

To view the set temperature press the star "    " key. To change the 
set temperature, hold the star key together with the up (raise temp) 
or down (lower temp) arrow key until the desired temperature is 
indicated on the LED display. 

(FIG 2)

  The temperature control is 

set at the factory to read in 1/10 degree C (centigrade) units. To 
change temperature units or display resolution see: Menu Level 
Functions (page 3).

Once the unit reaches the set temperature, allow the unit to cycle 
for 20 minutes at set point before temperature becomes fully stable.
NOTE: Upon each initial powering-up, the control may typically overshoot the set temp by 2 to 4 degrees, 
especially if the temperature set point is close to the operating ambient temperature. After equilibrium is achieved 
the control will hold set temperature within 1 unit degree.

Article processing times and temperature uniformity are largely dependent on load density and positioning. Load 
the incubator so that the air circulation within the incubator is not impaired.  Here are some general guidelines:

Leave a space between articles on a shelf. Stagger articles from those on lower shelves. 

Avoid placing articles or media against or within an inch of the walls, especially on the lower shelf. Heated air from 
the lower plenum openings, designed to travel up the side walls, can have a slightly elevated temperature from set 
point and the rest of the chamber.

Use of large solid trays or foil on shelves limits heat to any articles placed on shelves above.

Avoid extremely large (in quantity or size), or high-density loads. This will show by non-uniform processing and 
long or impossible "heat-through" times. To help determine a load's suitability, use the set-point recovery time (the 
time it takes for the temperature to recover to the original set temperature once the load is placed), as a guide. To 
reduce recovery time, reduce load proportionally. When possible, measure large loads or solution temperatures 
directly with an ancillary thermometer or probe. Probes can be inserted at the top port.

For best processing performance for a single item, adjust one shelf so that the article is centered in the incubator 
chamber.

Содержание 140E-1 Series

Страница 1: ...rd Electrical 230 volts 50 60 Hz 31x25x25 46x41x30 46x41x30 31x25x25 33x38x28 48x53x33 33x38x28 48x53x33 SPECIFICATIONS MODEL 10 140E 1 MODEL 12 140E 1 MODEL 10 180E 1 MODEL 12 180E 1 Common Unit Construction Exterior Powder Coated Steel Interior Aluminum Insulation Fiberglass Door 140E 1 Acrylic 180E 1 Steel Insulated Thermo control PID Microprocessor Heater Resistive Tubular Incoloy WATTS 120 23...

Страница 2: ...e up raise temp or down lower temp arrow key until the desired temperature is indicated on the LED display FIG 2 The temperature control is set at the factory to read in 1 10 degree C centigrade units To change temperature units or display resolution see Menu Level Functions page 3 Once the unit reaches the set temperature allow the unit to cycle for 20 minutes at set point before temperature beco...

Страница 3: ... 2 Change to whole degree or 1 10th degree display resolution in level 2 Run or Read temperature tracking or in level 3 Lock Temperature setting against inadvertent adjustment in level 1 Calibrate control temperature to an external standard in level 3 To access the control s function menu Press and hold both arrow keys for 3 seconds then release when the tune function prompt is displayed from with...

Страница 4: ... menu level 3 Hold star key then up arrow key to select ON Return to temp display or control will auto return in 60 seconds The control will track the temperature variation until CheK is turned off Recorded readings are retained until next CheK ON You can view readings at any time during or after tracking feature has been turned off But de powering the unit will reset CheK to OFF and rEAd to zero ...

Отзывы: