background image

Quincy Lab, Inc.

      

1925 N. Leamington Ave.       Chicago, Illinois  60639    1-800-482-HEAT (4328)

*

*

*

*

Quincy Lab, Inc.

View Setpoint

(PRESS)

Increase Setpoint

Decrease Setpoint

(3 sec.) Enter/Exit Menu Leve

ls

Model 

 12-140E  ncubator

POWER

DIGITAL M

ICROPROC

ESSOR

140 SERIE

S

Switch

opening

Quincy Lab, Inc.

PAGE 1

Issue Rev C 0310

Copyright Quincy Lab Inc. 2009

Model Series 140E-1 & 180E-1 (230 V   )

General Purpose Incubators

OPERATING MANUAL

I

Model 10-180E-1

DIGITAL M

ICROPROC

ESSOR

*

*

*

Quincy Lab, Inc.

View Setpoint

(PRESS)

Increase Setpoint

Decrease Se

tpoint

(3 sec.) Ente

r/Exit Menu Levels

Model 

 10-180E  ncubator

I

POWER

Switch

opening

140 SERIES

Model 12-140E-1

Interior Dimensions

INCHES W x H x D

(CM) W x H x D

Exterior Dimensions

INCHES W x H x D

(CM) W x H x D

12x10x10

18x16x12

18x16x12

12x10x10

13x15x11

19x21x13

13x15x11

19x21x11

Standard Electrical:   

230 volts, 50/60 Hz

31x25x25

46x41x30

46x41x30

31x25x25

33x38x28

48x53x33

33x38x28

48x53x33

SPECIFICATIONS

MODEL

   10-140E-1

MODEL

   12-140E-1

MODEL

   10-180E-1

MODEL

   12-180E-1 

Common Unit Construction

Exterior

:  Powder-Coated Steel                   

Interior:  

Aluminum

Insulation:

  Fiberglass   

 

     

Door:  

140E-1:Acrylic, 180E-1:Steel Insulated 

Thermo-control: 

 PID Microprocessor       

Heater: 

Resistive-Tubular Incoloy

     

WATTS

120

235

270

385

Cubic Foot Capacity

.7 ft

3

2.0 ft

3

.7 ft

3

2.0 ft

3

Common Unit Specifications

Operating Environment

:  

Indoor use, altitude to 6,500 ft. (2,000m) Installation Category II,

 

Pollution Degree 2, ambient temperature 10

o

C/50

o

F to 35

o

C/95

o

F, 

80% RH maximum.

Storage Temperature:

      -10

o

C/14

o

F to 70

o

C/158

o

F, 70% RH maximum.

Approvals:

Underwriter's Laboratory Listed, Laboratory Equipment, C/UL
United States/Canadian. E212550 (115VAC models only)

Compliance:

                     UL Standard 61010-1, IEC 61010-1, 2nd Edition.

Weight (lbs)

19 lbs

33 lbs

19 lbs

33 lbs

Temperature Range

A 2

o

C to 62

o

C

A 3

o

C to 94

o

C

Содержание 140E-1 Series

Страница 1: ...rd Electrical 230 volts 50 60 Hz 31x25x25 46x41x30 46x41x30 31x25x25 33x38x28 48x53x33 33x38x28 48x53x33 SPECIFICATIONS MODEL 10 140E 1 MODEL 12 140E 1 MODEL 10 180E 1 MODEL 12 180E 1 Common Unit Construction Exterior Powder Coated Steel Interior Aluminum Insulation Fiberglass Door 140E 1 Acrylic 180E 1 Steel Insulated Thermo control PID Microprocessor Heater Resistive Tubular Incoloy WATTS 120 23...

Страница 2: ...e up raise temp or down lower temp arrow key until the desired temperature is indicated on the LED display FIG 2 The temperature control is set at the factory to read in 1 10 degree C centigrade units To change temperature units or display resolution see Menu Level Functions page 3 Once the unit reaches the set temperature allow the unit to cycle for 20 minutes at set point before temperature beco...

Страница 3: ... 2 Change to whole degree or 1 10th degree display resolution in level 2 Run or Read temperature tracking or in level 3 Lock Temperature setting against inadvertent adjustment in level 1 Calibrate control temperature to an external standard in level 3 To access the control s function menu Press and hold both arrow keys for 3 seconds then release when the tune function prompt is displayed from with...

Страница 4: ... menu level 3 Hold star key then up arrow key to select ON Return to temp display or control will auto return in 60 seconds The control will track the temperature variation until CheK is turned off Recorded readings are retained until next CheK ON You can view readings at any time during or after tracking feature has been turned off But de powering the unit will reset CheK to OFF and rEAd to zero ...

Отзывы: