Operating Procedures
PyroMark Q96 ID User Manual 01/2016
5-1
5
Operating Procedures
This section describes how to operate the PyroMark Q96 ID.
Before proceeding, it is recommended that you familiarize
yourself with the features of the PyroMark Q96 ID by
referring to Sections 3.4 and 3.5.
5.1
Required items
The following items are needed to perform a run:
Operator’s computer loaded with PyroMark Q96
Software
PyroMark Q96 Vacuum Workstation
Sample preparation buffers
PyroMark Gold Q96 Reagents
PyroMark Q96 ID Cartridge
PyroMark Q96 Plate Low
High-purity water (Milli-Q
®
18.2 MΩ x cm or equivalent)
Powder-free gloves
Lint-free tissues
5.2
Starting the system
The PyroMark Q96 ID requires a warm-up time of 90
minutes for the CCD camera output to stabilize. To save
time, switch on the instrument before starting to prepare
samples.
5.2.1
Starting the PyroMark Q96 ID
Before starting the PyroMark Q96 ID, ensure that the mains
power cable is correctly inserted into the mains power
connector on the rear panel and connected to a grounded
mains power socket.
Turn the power switch to
I
(On). All indicator lights turn on
when power is connected. After 1–3 minutes, the “Busy” light
will turn off and the “Information” light will start flashing. The
instrument is now ready for a run.
Содержание PyroMark Q96 ID
Страница 1: ...Version 5 January 2016 Sample Assay Technologies PyroMark Q96 ID User Manual...
Страница 2: ......
Страница 8: ...Contents Contents 6 PyroMark Q96 ID User Manual 01 2016 This page intentionally left blank...
Страница 14: ...Safety Information 1 6 PyroMark Q96 ID User Manual 01 2016 This page intentionally left blank...
Страница 32: ...Installation Procedures 4 6 PyroMark Q96 ID User Manual 01 2016 This page intentionally left blank...
Страница 58: ...Operating Procedures 5 26 PyroMark Q96 ID User Manual 01 2016 This page intentionally left blank...
Страница 78: ...Maintenance 7 12 PyroMark Q96 ID User Manual 01 2016 This page intentionally left blank...
Страница 96: ...Appendix A A 8 PyroMark Q96 ID User Manual 01 2016 This page intentionally left blank...
Страница 106: ...Appendix C C 2 PyroMark Q96 ID User Manual 01 2016 This page intentionally left blank...
Страница 114: ...Appendix D D 8 PyroMark Q96 ID User Manual 01 2016 This page intentionally left blank...
Страница 122: ...Appendix E E 8 PyroMark Q96 ID User Manual 01 2016 This page intentionally left blank...
Страница 125: ......
Страница 126: ...1100446 01 2016 Sample Assay Technologies...