Hybrid Capture System Rotary Shaker 1 User Manual 09/2018
6.4
Replacing the fuse
Only use fuses of the same type and rating for the voltage in your location. Note: a spare
fuse is stored in the fuse drawer
1. Remove the power cord from the power source.
2. Gently pry open the fuse drawer at the rear of the shaker base.
3. Remove the fuse from the plastic clip in the fuse drawer
4. Place a new fuse in plastic clip in the fuse drawer.
5. Insert the fuse drawer back into the instrument.
Voltage
HCS Rotary Shaker 1
catalog number
Fuse
Amperage
Fuse
Type
120 V
6000-2110E
5 AMP
250 volt
5 x 20 mm
(0.25 x 1.25 in.)
Quick acting UL-listed fuse
230 V
6000-2240E
5 AMP
250 volt
5 x 20 mm
(0.25 x 1.25 in.)
Quick acting UL-listed fuse