Hybrid Capture System Rotary Shaker 1 User Manual 09/2018
25
3. Check all platform screws to make sure they are properly tightened
6.3
Detaching and replacing the shaker platform
In the event that the shaker platform needs to be removed (eg. to clean up spilled reagent that
cannot be properly cleaned with the platform in place), detach and replace the shaker platform
as follows:
1. Remove the 4 screws on the shaker platform that are covered when plates are loaded on
the platform.
2. Lift the shaker platform up off the shaker base.
3. Clean the shaker platform and shaker base with a dilute detergent solution. Make sure
the shaker platform is completely dry before proceeding to the next step.
4. Line up the 4 screw holes on the shaker base with the 4 holes on the shaker platform.
5. Secure the shaker platform to the base mount using the 4 screws that were removed
earlier.