Multi-Pointer
Service Manual
Page 21 of 26
State 3
State 4
Fig. 29
2. When the up and down laser point error is as shown in the Fig. 29, you can adjust
the power adjusting screw to calibrate (see Fig. 28). When the instrument is in the
state 3, you can loose the screw 4 and screw down 3. When the instrument is in the
state4, you can loose the screw 3 and screw down 4.
State 5
State 6
Fig. 30
3. When the up and down laser point error is as shown in the Fig. 30, you can adjust
the pendulum adjusting screw to calibrate (see Fig. 31). When the instrument is in
the state 5, you need to twist the pendulum adjusting screw clockwise using an
inner hexagon spanner. If in the state 6, you need to twist the pendulum adjusting
screw count-clockwise.
Fig. 31