5K-MNL-BIOGAS5000
BIOGAS 5000 gas analyzer
Page
46
9.0 Calibration
9.1 Calibration introduction
The BIOGAS 5000 gas analyzer is carefully calibrated at manufacture and when returned for
service. However, it is sometimes desirable to be able to carry out a calibration process
between services.
This section outlines the correct procedures to enable the site engineer to field calibrate the
gas analyzer.
Note:
This does not replace the factory service and calibration. If this calibration is
completed incorrectly it may decrease the accuracy of the gas analyzer.
CH
4,
CO
2
and O
2
can be measured by BIOGAS 5000 gas analyzer as standard; these channels
can be user calibrated. The analyzers have other gas channel options that are specified at
manufacture; these too can be calibrated. This section will describe in detail how to calibrate
the three standard gas channels plus the CO channel.
The BIOGAS 5000 instrument can have a H
2
compensated CO channel. This option requires
that H
2
be used in the calibration process and is also set out within this section.
For the other gas channel options contact LANDTEC for advice.
Two important terms that are used within this section are ‘Zero’ and ‘Span’.
Zero:
The point at which the gas analyzer is calibrated when there is none of the target
gases present.
Span:
The point at which the gas analyzer is calibrated when a known quantity of the
target gas is present.
9.2 Frequency of calibration – best practice
The BIOGAS 5000 gas analyzers can be checked against a known concentration of gas, to
give confidence that the analyzer is operating as expected at the time and conditions in which
it is being used.
It is recommended that the instrument be regularly serviced and calibrated by LANDTEC in
accordance with the due date on the instrument.
When defining the frequency of user calibration, the following are factors to be considered:
The frequency of use of the analyzer. (daily?/monthly?)
The level of confidence and accuracy required for readings to be taken.
Historical user calibration data.
Site specific requirements or conditions.
Historical understanding of expected readings on site.
Zeroing of the gas analyzer should be undertaken at the start of each day’s monitoring.
Use historical data to drive your frequency of calibration.
If there is no historical data a good starting point for a daily monitoring round is performing a
calibration once every week or every other week.
Содержание BIOGAS 5000
Страница 44: ...5K MNL BIOGAS5000 BIOGAS 5000 gas analyzer Page 44 8 9 Taking gas and flow measurement...
Страница 45: ...BIOGAS 5000 Gas Analyzer 5K MNL BIOGAS5000 Page 45...
Страница 53: ...BIOGAS 5000 Gas Analyzer 5K MNL BIOGAS5000 Page 53...
Страница 61: ...BIOGAS 5000 Gas Analyzer 5K MNL BIOGAS5000 Page 61...
Страница 82: ...5K MNL BIOGAS5000 BIOGAS 5000 gas analyzer Page 82 Dr Roger Riley...
Страница 103: ...BIOGAS 5000 Gas Analyzer 5K MNL BIOGAS5000 Page 103 Dr Roger Riley...