Pyxis ST-601
Operation Manual
14
Figure 10. Connected to a ST-601 probe and information page
Figure 11. Calibration page
4.2.1
Calibration
Click
Calibration
to launch the calibration page (Figure 11). Place the probe in deionized water and click
Zero Calibration
to carry out the zero-point calibration.
Place the probe in a known ClO2 calibration standard and tap
Slope Calibration
to carry out the slope
calibration. Enter the ClO2 concentration in the dialog window as in Figure 12. For the best result, the
ClO2 standard concentration should be in the range of 1000ppm to 3000ppm for ST-601.