background image

DW-2100P User 

[email protected] 

|  +1 (866) 203-8397 

DW-2100P Features

 

Pyxis Lab’s advanced research and development sensor technologies to achieve highly accurate and stable 

measurement of Turbidity, Free Chlorine, pH and Temperature. 

Pyxis LT-739 (EPA) and LT-739B (ISO) ultra-low resolution turbidity sensors offer a detection light source 

using warm white LED in 90-degree surface scatter format in accordance with USEPA 180.1 standards or 

Infra-Red 860nm in accordance with ISO-7027.1 standards. The turbidity sensors are mounted in the 

unique Pyxis FR-200 dual sensor flow reservoir enabling the highest resolution possible of 0.001NTU 

with unmatched stability. The LT-739 and LT-739B offers simple calibration via the Pyxis 

L

T-

SOLID-

CAL Solid State Turbidity Calibration Kit (no Formazin needed) or the Pyxis L-CAL Portable Liquid 

Turbidity Calibration Kit (uses 500mL per calibration). 

Pyxis ST-765SS-FCL three-parameter composite sensor is used for the measurement residual free chlorine, 

pH and temperature in compliance with USEPA 334.0 and ISO-7393 guidelines.  The sensors advanced PCB 

offers built-in temperature and pH parameter compensation (up to pH 9.0) algorithms eliminating the need 

for a supplemental pH sensor and controller. Unique Bare-Gold electrode technology for residual chlorine 

measurement eliminates membranes and electrode solution replenishment commonly associated with 

conventional sensors. The ST-765SS-FCL has a uniquely designed flat bubble pH electrode design for 

reduced fouling potential. Reduce your maintenance and cost versus colorimetric chlorine measurement or 

conventional electrochemical sensors by utilizing Pyxis replaceable Electrode Head (EH-765) for this sensor 

allowing for years of reliable service. The ST-765SS-FCL may be calibrated in-situ after cleaning via DPD Free 

Chlorine wet chemistry test measurement of active sample. 

Pyxis FR-200 Dual-Sensor flow reservoir provides sample calming for dissipation of air-bubbles and 

settling of suspended solids, foam or other impurities commonly observed in drinking water influent 

with a recommended flow range of 600–1,000mL/minute. This unique flow reservoir design results in 

the highest level of turbidity resolution on the market and greatly extends the maintenance cycle of the 

sensor while providing a large buffer capacity to mitigate pressure fluctuations. The minimum inlet pressure 

of FR-200 flow reservoir is only 7.5 psi (0.05mpa) making it highly suitable for the end of pipe networks and 

secondary supply influent sampling. 

Both sensors offer a self-cleaning flat electrode design making them easy to maintain and clean. 

Simple sensor removal and replacement. Both sensors are connected to the display/data logger via RS-485 

Modbus (RTU) allowing for integrated sensor calibration interface and diagnostics within the display touch 

screen. 

Convenient and simple to install Back-Panel (DW-2100P) for rapid and easy installation. Truly a plumb 

and power to go platform with intense factory setup, testing and sensor calibration prior to shipment. 

Touch screen display/data logger interface with sensor calibration integrated. Display/data logger offers 2x 

4-20mA I/O as well as RS-485 and TCP-IP with remote diagnosis and parameter adjustment. Pyxis

 

CloudLink

TM

 4G Gateway version available.

Содержание DW-2100P

Страница 1: ...Pyxis Lab Inc 1729 Majestic Drive Suite OS Lafayette Colorado 80026 www pyxis lab com...

Страница 2: ...Parameter Inline Analyzers for Drinking Water User Manual January 31 2022 Rev 1 03 Pyxis Lab Inc 1729 Majestic Dr Suite 5 Lafayette CO 80026 USA www pyxis lab com 2017 Pyxis Lab Inc Pyxis Lab Proprie...

Страница 3: ...e at the time of delivery All service interventions are to be reviewed and authorized as correct and complete at the completion of the service by a customer representative or des ignate Pyxis warrants...

Страница 4: ...needs special emphasis Warning Label Please read all labels and marks attached to the instrument Failure to follow the instructions on these safety labels may result in personal injury or damage to th...

Страница 5: ...s 18 6 5 1 Cleaning Parameters Setting Using FRP 200 Optional Automatic Cleaning Module 18 6 5 2 Alarm Parameters Setting 19 6 5 3 Diagnostic Parameters 20 6 5 4 4 20mA Output Parameters Setting 21 6...

Страница 6: ...6 260VAC 50 60 Hz 10AFuse 200 W Output 3 x 4 20 mA RS 485 Modbus RTU Modbus TCP Input 3 x 4 20 mA RS 485 Modbus RTU USB 1 x USB host for data downloading and screen upgrade Internet RJ 45 socket Modbu...

Страница 7: ...g in G System Layout and Features The DW 2100P series are multi parameter inline water analyzers specifically designed as a Turn Key monitoring solution for clean water applications including drinking...

Страница 8: ...e and cost versus colorimetric chlorine measurement or conventional electrochemical sensors by utilizing Pyxis replaceable Electrode Head EH 765 for this sensor allowing for years of reliable service...

Страница 9: ...DW 2100P User service pyxis lab com 1 866 203 8397 9 Dimension and Mounting DW 2100P Series Panel Dimensions mm DW 2100P Series UC 100A Controller Dimensions mm...

Страница 10: ...00P User service pyxis lab com 1 866 203 8397 10 LT 739 LT 739B Turbidity Sensor ST 765SS FCL Free Chlorine pH Sensor DW 2100P Series FR 200 Flow Reservoir Dimensions mm DW2100P Series Sensor Dimensio...

Страница 11: ...cted to a discharge drain via gravity flow Wall Mount Space The DW 2100P analyzer panel size is roughly 790H x 450W x 235D mm in dimension Please accommodate sufficient space for mounting Wall Mount W...

Страница 12: ...DW 2100P User service pyxis lab com 1 866 203 8397 12 Figure 2 FR 200 Connections...

Страница 13: ...866 203 8397 13 5 3 Terminal Wiring The DW 2100P analyzer has universal AC power supply equipment allowing users simply to plug the power supply into a 100 240V AC 50 60Hz power outlet for normal oper...

Страница 14: ...al Diagram Warning The process of electrical connection to contact the 220V single phase power supply should be operated by personnel with an electrician s license Failure to operate according to the...

Страница 15: ...log into the system Figure 4 Main Screen 6 2 User Login After powering on the system log in with the username and password to be able to change system settings Click the User Login button select the...

Страница 16: ...l be displayed in real time See a functional overview of each section of this screen highlighted below numbers 1 4 Figure 6 Real time monitoring screen Section Number Functional Overview 1 The blue ar...

Страница 17: ...he display value range of Y axis for each measurement index curve Click the outter area of the screen to save and exit the setting screen after modifications are made Figure 7 Curve Range Setting 6 4...

Страница 18: ...gure 9 Parameter Settings 6 5 1 Cleaning Parameters Setting Using FRP 200 Optional Automatic Cleaning Module When entering the cleaning control parameter setting interface for the first time a reminde...

Страница 19: ...ting with Optional FRP 200 Module 6 5 2 Alarm Parameters Setting Users can set the upper and lower alarm limits Click Alarm Parameters to enter the alarm parameter settings When the measured sensor va...

Страница 20: ...the diagnosis page the raw data measured by the sensor is displayed To help troubleshooting possible issues with the sensor please save an image of this data when the sensor is placed in a clean wate...

Страница 21: ...sensor range NOTE The closer the value is set to the measurement value the more accurate the data It is recommended to set according to the range of the sensor Figure 14 4 20mA Output Setting 6 5 5 C...

Страница 22: ...16 Sensor Calibration 6 6 1 pH Calibration The pH function is thoroughly calibrated at the Pyxis Lab factory After checking with a pH standard buffer solution if the sensor value has shifted then the...

Страница 23: ...ion A dialog box will pop up to confirm whether to perform this operation click OK if the calibration operation is confirmed if the calibration is successful the dialog box will show calibration succe...

Страница 24: ...pop up to confirm whether to perform this operation Click OK if the calibration operation is confirmed and the dialog box will show Calibration Successful if the calibration is successful After succe...

Страница 25: ...m this operation If the calibration operation is confirmed click OK and if the calibration is successful the dialog box will show Calibration Success NOTE As an alternative a standard residual chlorin...

Страница 26: ...uccessful the dialog box will show Calibration Success The sensor is now blank calibrated to the known zero calibration solution Slope Calibration Procedure Use a portable or laboratory colorimeter to...

Страница 27: ...orm a diagnostic check on the sensor to determine if the sensor needs to be calibrated To complete this task the user must drain the FR 200 flow reservoir and then wipe the FR 200 flow cell and sensor...

Страница 28: ...Calibration while using Bubble Free DI Water Figure 23 Awaiting execution Screen of Turbidity Calibration Mid Range Calibration using the FR 200 Flow Reservoir After the low range calibration has bee...

Страница 29: ...ing items should be checked 1 Ensure your source of Deionized water is not contaminated with turbidity 2 Ensure your turbidity calibration standard solutions have not been contaminated 3 Ensure the LT...

Страница 30: ...libration L CAL Liquid Turbidity Calibration Kit P N 53247 Turbidity Calibration Principals Considerations The precision resolution and the low detection limit of the LT 73X Series sensors are not aff...

Страница 31: ...DW 2100P User service pyxis lab com 1 866 203 8397 31...

Страница 32: ...consistent and reproducible method to calibrate Pyxis inline LT 739 series turbidity sensors overcoming the challenges associated with the liquid standard calibration LT SOLID CAL Specifications Item...

Страница 33: ...e data of pH residual chlorine sensor and turbidity sensor If a user error is made during calibration and other operations you may restore the factory settings of the sensor through the restore functi...

Страница 34: ...then enter the alarm number in the pop up box to query that alarm Figure 26 Alarm Data Query Screen The Delete button in the lower left corner will delete all alarm records After clicking delete you...

Страница 35: ...and Next to view historical data records The data record can save up to 100 000 data entries New data will overwrite the previously saved data after recording 100 000 data entries The user can click...

Страница 36: ...he time range of the data to be exported in the query area Click on the Data Export to download the data to the USB disk The data quantity will be shown as a positive number if data export is successf...

Страница 37: ...DW 2100P User service pyxis lab com 1 866 203 8397 37 Figure 31 History Curve Screen...

Страница 38: ...DW 2100P User service pyxis lab com 1 866 203 8397 38 Figure 32 Y axis Range Setting Figure 33 Button Function Review Figure 34 Button Function Review...

Страница 39: ...ce pyxis lab com 1 866 203 8397 39 Figure 35 Time Setting Screen 6 11 User Management Click the User Management button on the menu bar and then you can select Login Logout and Manage operations Figure...

Страница 40: ...rs can set their own user name and password and select the user group they belong to Only users in the administrator group can set parameters such as calibration Modify Figure 37 User Management Scree...

Страница 41: ...al Chlorine Lower Limit Alarm 12 unit Read Only 0 Normal 1 Alarm 10 Residual Chlorine pH Sensor Communication Abnormalities 13 unit Read Only 0 Normal 1 Alarm 11 Turbidity Sensor Comm Abnormalities 14...

Страница 42: ...ce may vary depending on the installation conditions and usage Please refer to the table below as a general recommended maintenance schedule Minimal operator intervention is required during normal ope...

Страница 43: ...ormation Only Turbidity Lower Limit Alarm Turbidity below the Alarm Setting Information Only Residual Chlorine Upper Limit Alarm Residual Chlorine above the Alarm Setting Information Only Residual Chl...

Страница 44: ...til the front end of the black electrode is completely unscrewed as shown in Figure 39 2 4 Pull out the electrode head as shown in Figure 39 3 5 Loosen the electrode plug connector and remove the elec...

Страница 45: ...DW 2100P User service pyxis lab com 1 866 203 8397 45 9 Order Information Table 5 Order Information...

Отзывы: