LT-739 Calibration using
L
T-
SOLID-
CAL Solid State Turbidity Calibration Kit
Pyxis Lab has developed reusable solid-state turbidity calibration kits for rapid calibration of the Pyxis LT-739
Series inline ultra-low turbidity sensors. The LT-739 Series ultra-low turbidity sensors are factory calibrated
on Formazin liquid standards. The
L
T-
SOLID-
CAL calibration kits represent an alternative to
Formazin calibration for routine field calibration only. The Pyxis solid-state calibration kits provide a
consistent and reproducible method to calibrate Pyxis inline LT-739 series turbidity sensors,
overcoming the challenges associated with the liquid standard calibration.
L
T-
SOLID-
CAL Specifications
Item
L
T-
SOLID-
CAL 739
L
T-
SOLID-
CAL 739B
P/N
53228
53238
Sensor Calibrated
LT-739
LT-739B
Quantity Adapters Included
3
Turbidities Included
0.10
±
0.05NTU
8.0
±
0.8NTU
25.0
±
2.5NTU
The sensor should be removed from process flow, cleaned gently with a soft towel, inserted into
the appropriate
L
T-
SOLID-
CAL adapter and wireless calibrated via the uPyxis Mobile or Desktop APP
to the designated NTU value assigned to the
L
T-
SOLID-
CAL adapter. Each
L
T-
SOLID-
CAL kit comes
complete with Low, Mid and High Range calibration adapters specific to that LT-73X Series sensor.
Содержание DW-2100P
Страница 1: ...Pyxis Lab Inc 1729 Majestic Drive Suite OS Lafayette Colorado 80026 www pyxis lab com...
Страница 12: ...DW 2100P User service pyxis lab com 1 866 203 8397 12 Figure 2 FR 200 Connections...
Страница 31: ...DW 2100P User service pyxis lab com 1 866 203 8397 31...
Страница 37: ...DW 2100P User service pyxis lab com 1 866 203 8397 37 Figure 31 History Curve Screen...
Страница 45: ...DW 2100P User service pyxis lab com 1 866 203 8397 45 9 Order Information Table 5 Order Information...