6.4
Reading Screen
The
READING
screen displays the current volume of liquid remaining and liquid level over time.
Figure 8.
6.5
Settings Screen
The sensor measures the distance between the liquid surface in the tank and the bottom sensor surface.
Converting this measured distance to other parameters such as the tank level, the remaining liquid volume,
or the consumed liquid, requires the tank dimensional and volume capacity information. Common vertical
tanks have a uniform horizontal cross section.
*
NOTE
*
For horizontal or other tanks that have a non-uniform cross section, please contact
Pyxis Technical Support team (
) for assistance.
To convert the measured distance to volumetric information, the user must enter three parameters via the
uPyxis®
Mobile App’s
Setting
screen:
•
Tank Volume
(rated volume capacity of the tank)
•
Installation Height
(from the bottom of the tank to the bottom of the LS-200)
•
Max Level Height
(from the bottom of the tank to the liquid surface when filled to the rated capacity)
*
NOTE
*
Max Level Height
must be at least 4 inches less than
Installation Height
Figure 9 illustrates these parameters for the tank.
In the
SETTING
screen, the you can also set the
Device Name
, the
Sampling Mode
,
LCD Display Value
,
LCD
Screen Off Time (in seconds)
, and the
Display Unit
. To save the settings, press
APPLY SETTINGS
. You may
also create or change a password for the LS-200 by pressing
MANAGE PASSWORD
.
LS-200 User Manual
| +1 (866) 203-8397
8