pH High Calibration
4.2.3
4-20mA Span
The default oxidizer 4-20mA span for ST-765SS Series probes is as follows.
ST-765SS-FCL
4mA=0ppm Free Chlorine / 20mA = 5ppm Free Chlorine
ST-765SS-CLO
4mA=0ppm Chlorine Dioxide / 20mA = 5ppm Chlorine Dioxide
Tap
4-20mA
SPAN
to change the free chlorine or chlorine dioxide value corresponding to the 20mA output
to a lower value as seen in
Figure 20
.
*NOTE*
The 4-20mA Span feature allows users to REDUCE the upper
20mA output scale only. You cannot INCREASE the upper limit of the sensor beyond the range of the
sensor.