| +1 (866) 203-8397
19
Figure 1
8
Slope Calibration
4.2.2
pH Calibration
Remove and place the sensor in a low pH (ie. 4.0) calibration standard solution and tap
pH LOW
CALIBRATION
in the uPyxis app. Measure for 1 minute and wait for the measurement result to stabilize
before performing calibration, the low pH calibration standard value range acceptable for this step is
1.00-6.00 pH.
Place the sensor into the pH 7.0 calibration standard solution and tap
pH 7 CALIBRATION
in the uPyxis app.
Measure for 1 minute and wait for the measurement result to stabilize before performing calibration. Place
the sensor in a high pH (ie. 10.0) calibration standard solution and tap
pH HIGH CALIBRATION
in the uPyxis
app. Measure for 1 minute and wait for the measurement result to stabilize before performing
calibration, the high pH calibration standard value range acceptable for this step is 8.00-13.00 pH.