Pyxis 53201 Скачать руководство пользователя страница 4

1

Introduction

The Pyxis ST-730 series is a variety of pre-scaled inline probes all measuring turbidity in water using a white
LED as the excitation light source and by measuring the scattered light at a 90-degree angle with respect to
the excitation beam. The fluidic and optical arrangement of the ST-730 series probe is designed to overcome
many shortcomings associated with other inline turbidimeters. It can be easily inserted into the custom-
made tee with a compression fitting port designed to ensure correct vertical positioning of the ST-730 series
probe in the fluid stream. The ST-730 series probe custom mounting tee has two ¾ inch female NPT ports
for plumbing into an existing ¾ inch sample water line. The ST-730 series probe can be connected to any
device that accepts an isolated or non-isolated 4-20mA input or RS-485 Modbus input. The ST-730 series
probe has a short fluidic channel that can be easily cleaned and calibrated using the MA-WB Bluetooth
adapter and

uPyxis®

Mobile or Desktop App.

2

Specifications

Specifications

ST-730

ST-730B

ST-730SS

ST-731

ST-735

P/N

53201

53202

50626

53505

53204

Range (NTU)

0-100

0-1,000

0-100

0-10

0-10,000

Resolution (NTU)

0.1

1

0.1

0.05

10

Accuracy

±2% of reading

Method

Nephelometric, with white LED and IR LED (860 nm) light sources

Calibration

Two-point calibration against standard solution

Outputs

4-20mA Analog Output, RS-485 Digital Output with Modbus protocol

Installation

*

Custom tee assembly (P/N: ST-001) with 3/4” female socket & NPT threaded ports

Cable Length

5 ft, with IP67 connectors. 30 ft and 60 ft extension cables available

Power Supply

22-26 VDC, 1 W

Dimension inch (mm)

Length: 6.8 (172.7), Body Diameter: 1.44 (36.6)

Weight lbs(g)

0.37 (170)

0.37 (170)

2.5 (1130)

0.37 (170)

0.37 (170)

Material

CPVC

CPVC

304 Stainless Steel

CPVC

CPVC

Operational

Tempera-

ture °F (°C)

40-120

(4-49)

40-120

(4-49)

32-104 (0-40)

40-120

(4-49)

40-120

(4-49)

Storage Temperature °F
(°C)

20-150 (-7-66)

Pressure psi (MPa)

Up to 100

(0.7)

Up to 100

(0.7)

Up to 290 (2.0) at 149 °F

(65 °C)

Up to 100

(0.7)

Up to 100

(0.7)

Enclosure Rating

IP66

Regulation

CE

*

The ST-730SS probe does not come with the custom tee assembly (P/N: ST-001). It has 3/4” female NPT threaded ports on the probe
itself.

3

Unpacking Instrument

Remove the instrument and accessories from the shipping container and inspect each item for any damage
that may have occurred during shipping. Verify that all accessory items are included. If any item is missing
or damaged, please contact Pyxis Lab Customer Service at

[email protected]

.

ST-730 Series User Manual

[email protected]

| +1 (866) 203-8397

2

Содержание 53201

Страница 1: ...0 Series Inline Turbidity Probes User Manual March 9 2020 Rev 1 33 Pyxis Lab Inc 1729 Majestic Dr Suite 5 Lafayette CO 80026 USA www pyxis lab com 2017 Pyxis Lab Inc Pyxis Lab Proprietary and Confiden...

Страница 2: ...Download uPyxis Mobile App 7 5 2 Connecting to uPyxis Mobile App 7 5 3 Calibration Screen and Reading 8 5 4 Diagnosis Screen 9 5 5 Device Info Screen 9 6 Setup and Calibration with uPyxis Desktop App...

Страница 3: ...very All service interventions are to be reviewed and authorized as correct and complete at the completion of the service by a customer representative or des ignate Pyxis warrants these services for 3...

Страница 4: ...1 1 0 1 0 05 10 Accuracy 2 of reading Method Nephelometric with white LED and IR LED 860 nm light sources Calibration Two point calibration against standard solution Outputs 4 20mA Analog Output RS 48...

Страница 5: ...Tee O ring and Nut P N ST 001 7 Pin Female Adapter Flying Leads Cable 2 ft P N MA 1100 User Manual available online at www pyxis lab com support html 3 2 Optional Accessories Figure 1 ST 730 Series Us...

Страница 6: ...the ST 730 series probe into the ST 001 Tee Assembly follow the steps below 1 Insert the provided O ring into the O ring groove on the tee 2 Insert the ST 730 series probe into the tee 3 Tighten the t...

Страница 7: ...ustom tee assembly The ST 730SS probe also has four 1 4 28 tapped mounting holes on the base of the sensor Figure 4 Dimension of the ST 730SS mm NOTE It is recommended to install the ST 730 series pro...

Страница 8: ...T 730 series probe to a controller Wire Color Designation Red 24V Black 24V Power ground White 4 20mA Green 4 20mA Blue RS 485 A Yellow RS 485 B Clear Shield earth ground Internally connected to the p...

Страница 9: ...y Figure 7 5 2 Connecting to uPyxis Mobile App Turn on Bluetooth on your mobile phone Do not pair the phone Bluetooth to the ST 730 series probe Open uPyxis Mobile App Once the app is open the app wil...

Страница 10: ...efault to the Calibration screen From the Calibration screen you can perform calibrations by pressing on Zero Calibration Slope Calibration and 4 20mA Span Follow the screen instructions for each cali...

Страница 11: ...well as Export Upload This feature may be used for technical support when communicating with service pyxis lab com Figure 10 5 5 Device Info Screen From the Device Info screen You can name the Device...

Страница 12: ...ver for the USB Bluetooth adapter MA NEB the USB RS485 adapter MA 485 and the main uPyxis Desktop application Double click the uPyxis Setup exe file to install Figure 12 Click Install to start the ins...

Страница 13: ...ormation description for Device Name and Product Name then click Set to save Figure 14 6 4 Calibration Screen To calibrate the device click on Calibration On the Calibration screen there are three cal...

Страница 14: ...instruction on how to calibrate then click Slope Calibration Figure 17 6 4 3 4 20mA Span To perform 4 20mA Span click on 4 20mA Span Then follow the instructions provided to alter the 4 20mA output sp...

Страница 15: ...es probe is configured as a Modbus slave device In addition to the NTU value many op erational parameters including warning and error messages are available via a Modbus RTU connection Contact Pyxis L...

Страница 16: ...nnel These components and more come with a Pyxis Lab Inline Probe Cleaning Solution Kit P N SER 01 which can be purchased at our online Estore Catalog https pyxis lab com product st series probe clean...

Страница 17: ...qualified turbidity standard After properly cleaning the ST 730 series sensor carry out the zero point and slope calibration using the qualified turbidity standard Pyxis Lab Turbidity Standards can be...

Отзывы: