background image

10.5

Setting Screen

From the

Setting

screen, the user can set the

Power off time

and

Screen off time

in seconds.

Figure 24.

11

Device Maintenance and Precaution

11.1

Maintenance Best Practices and Quick Tips

For greatly increased working life and ease of use of the SP-380P follow the list of maintenance best practices
and quick tips below:

• Rinse the sample cell at least three times with the desired solution before turning on.

• After a successful calibration, the unit does not automatically return to the measurement mode. If

Exit is held down too long the unit will power down rather than returning to the measurement mode.

• After returning to measurement mode after calibration, rinse several times with the first sample. The

unit will continue to read the sample values without any further key presses if it has not powered off.
If there are no key presses for 20 seconds the screen will darken, and after another 20 seconds without
key activity will power down. The press of any key while the screen is dark will reset the timer and
the screen will re-light. (This press will not step along the calibration process; the next press needed
will still be required in the sequence).

• Always rinse the unit with clean water after use and dry by clean tissue or paper towel.

11.2

Methods to Cleaning the SP-380P

A light deposit on quartz glass inside the conductivity cell can be cleaned by a Q-tip. Aged heavy deposi-
tion, especially iron oxide deposited, can be removed using a cleaning solution that is capable of removing
iron, such as the Pyxis Handheld Device Cleaning Solution Kit (P/N: SER-02) available from Pyxis online eS-
tore/Catalog

https://pyxis-lab.com/product/handheld-device-cleaning-kit/

.

SP-380P User Manual

[email protected]

| +1 (866) 203-8397

16

Содержание 50402

Страница 1: ......

Страница 2: ...Handheld Fluorometer User Manual October 20 2020 Rev 2 00 Pyxis Lab Inc 1729 Majestic Dr Suite 5 Lafayette CO 80026 USA www pyxis lab com 2017 Pyxis Lab Inc Pyxis Lab Proprietary and Confidential...

Страница 3: ...libration Two Point with Zero 7 8 Device Information and Diagnosis 8 8 1 Sample Cell Cleanliness Check 8 8 2 Bluetooth Connection to Devices 9 8 3 Factory Reset 9 9 Use with uPyxis Mobile App 11 9 1 D...

Страница 4: ...elivery All service interventions are to be reviewed and authorized as correct and complete at the completion of the service by a customer representative or des ignate Pyxis warrants these services fo...

Страница 5: ...iated with the cuvette are eliminated Large color graphic screen that can be read in direct sunlight 2 Specifications Table 1 SP 380P Specifications Item Specification Part Number P N 50402 Fluorescen...

Страница 6: ...s 100 ppb PTSA Calibration Standard 500 mL 21001 Pyxis 200 ppb PTSA Calibration Standard 500 mL 21031 Pyxis 10 ppm Fluorescent Polymer Calibration Standard 500 mL 21054 Pyxis 20 ppm Fluorescent Polyme...

Страница 7: ...and negative terminal signs in the compartment bottom snap four new AA alkaline batteries firmly into the battery holder 4 Replace the battery compartment cover and ensure that the sealing O ring is l...

Страница 8: ...e the sample water to rinse the sample cell at least three times Allow 5 10 seconds for the SP 380P to stabilize Figure 2 For a sample containing 10 ppm Fluorescent Polymer the measured Fluorescent Po...

Страница 9: ...the zero blank calibration 6 If the calibration succeeds a checkmark and instructions for the slope calibration will appear Figure 4 7 Rinse the sample cell three times with the desired Fluorescent Po...

Страница 10: ...zero point 6 If the calibration succeeds a checkmark and instructions for the slope calibration will appear Figure 7 7 Rinse the sample cell three times with the desired PTSA standard Fill the sample...

Страница 11: ...n HST Heavy fouling will pre vent the light from reaching the sensor resulting in inaccurate readings It is suggested that the SP 380P be checked for fouling and cleaned on a monthly basis Heavily con...

Страница 12: ...evices follow the steps below 1 Power on the SP 380P by pressing 2 Press Info to launch the DEVICE INFORMATION screen 3 Press Diagnosis or to launch the SYSTEM DIAGNOSIS screen 4 Allow5 10secondsforth...

Страница 13: ...igure 15 appear and the user can choose one of three options a Press to start the factory reset or b Press Cancel to return to the HELP screen or c Press Exit to abandon the factory reset entirely 7 A...

Страница 14: ...e phone Do not pair the phone Bluetooth to the SP 380P Open uPyxis Mobile App Ensure that the SP 380P is discoverable by following the steps in the Bluetooth Connection to Devices section Once the app...

Страница 15: ...well as update the firmware of the SP 380P by pressing Check Update If a firmware update is available press Get Firmware Once the new firmware is downloaded press Upgrade Firmware NOTE The firmware up...

Страница 16: ...Net Framework 4 5 if not previously installed on the PC the USB driver for the USB Bluetooth adapter MA NEB the USB RS485 adapter MA 485 and the main uPyxis Desktop application Double click the uPyxis...

Страница 17: ...Desktop App will default to the System screen From the System screen users can upgrade the firmware by selecting an appropriate firmware file contact service pyxis lab com for these firmware files and...

Страница 18: ...og List and selecting the desired datalog these are separated by month The SP 380P will then populate any relevant log event from the selected datalog which can be viewed in more detail by clicking Re...

Страница 19: ...hout any further key presses if it has not powered off If there are no key presses for 20 seconds the screen will darken and after another 20 seconds without key activity will power down The press of...

Страница 20: ...onfirm that the SP 380P is clean Repeat the process as needed until the Cleanliness Check shows Clean 11 3 Storage Do not expose the SP 380P to an extreme high or low temperature condition such as lea...

Отзывы: