background image

8

Device Information and Diagnosis

The

DEVICE INFORMATION

screen is launched when

Info

(

) is pressed in the measurement mode. This

screen contains the device serial number, software version, and hardware version (Figure 9). The battery
life as a percentage and the MAC addresses for main module also shown.

Press

Diagnosis

(

or

) to launch the

SYSTEM DIAGNOSIS

screen where raw measurement data

are displayed (Figure 10). The information has no use for normal operation, but instead is used for device
troubleshooting. Provide an image of both the

DEVICE INFORMATION

screen and the

SYSTEM DIAGNOSIS

screen when you contact Pyxis (

[email protected]

for troubleshooting your device or call +1 (866)

203-8397.

Figure 9.

Figure 10.

8.1

Sample Cell Cleanliness Check

The SP-380P is designed to provide reliable and accurate measurement on HST. Heavy fouling will pre-
vent the light from reaching the sensor, resulting in inaccurate readings. It is suggested that the SP-380P
be checked for fouling and cleaned on a monthly basis. Heavily contaminated waters may require more
frequent cleanings. Cleaner water sources with less contamination may not require cleaning for several
months. The SP-380P is designed to carry out a Cleanliness Check as described below:

1. Power on the SP-380P by pressing

.

2. Press

Info

(

) to launch the

DEVICE INFORMATION

screen.

3. Press

Diagnosis

(

or

) to launch the

SYSTEM DIAGNOSIS

screen.

4. Allow 5–10 seconds for the message in the top-right corner of the display change from

Starting BTLE...

to

BTLE Started

5. Press

Cleanliness

(

). An instruction prompt appears to ask the user to put DI water into the

sample cell (Figure 11).

6. Pour DI water into the sample cell.

7. Press

Confirm

(

,

, or

). The instruction prompt will disappear and the SP-380P displays

a countdown toward the bottom of the display.

SP-380P User Manual

[email protected]

| +1 (866) 203-8397

8

Содержание 50402

Страница 1: ......

Страница 2: ...Handheld Fluorometer User Manual October 20 2020 Rev 2 00 Pyxis Lab Inc 1729 Majestic Dr Suite 5 Lafayette CO 80026 USA www pyxis lab com 2017 Pyxis Lab Inc Pyxis Lab Proprietary and Confidential...

Страница 3: ...libration Two Point with Zero 7 8 Device Information and Diagnosis 8 8 1 Sample Cell Cleanliness Check 8 8 2 Bluetooth Connection to Devices 9 8 3 Factory Reset 9 9 Use with uPyxis Mobile App 11 9 1 D...

Страница 4: ...elivery All service interventions are to be reviewed and authorized as correct and complete at the completion of the service by a customer representative or des ignate Pyxis warrants these services fo...

Страница 5: ...iated with the cuvette are eliminated Large color graphic screen that can be read in direct sunlight 2 Specifications Table 1 SP 380P Specifications Item Specification Part Number P N 50402 Fluorescen...

Страница 6: ...s 100 ppb PTSA Calibration Standard 500 mL 21001 Pyxis 200 ppb PTSA Calibration Standard 500 mL 21031 Pyxis 10 ppm Fluorescent Polymer Calibration Standard 500 mL 21054 Pyxis 20 ppm Fluorescent Polyme...

Страница 7: ...and negative terminal signs in the compartment bottom snap four new AA alkaline batteries firmly into the battery holder 4 Replace the battery compartment cover and ensure that the sealing O ring is l...

Страница 8: ...e the sample water to rinse the sample cell at least three times Allow 5 10 seconds for the SP 380P to stabilize Figure 2 For a sample containing 10 ppm Fluorescent Polymer the measured Fluorescent Po...

Страница 9: ...the zero blank calibration 6 If the calibration succeeds a checkmark and instructions for the slope calibration will appear Figure 4 7 Rinse the sample cell three times with the desired Fluorescent Po...

Страница 10: ...zero point 6 If the calibration succeeds a checkmark and instructions for the slope calibration will appear Figure 7 7 Rinse the sample cell three times with the desired PTSA standard Fill the sample...

Страница 11: ...n HST Heavy fouling will pre vent the light from reaching the sensor resulting in inaccurate readings It is suggested that the SP 380P be checked for fouling and cleaned on a monthly basis Heavily con...

Страница 12: ...evices follow the steps below 1 Power on the SP 380P by pressing 2 Press Info to launch the DEVICE INFORMATION screen 3 Press Diagnosis or to launch the SYSTEM DIAGNOSIS screen 4 Allow5 10secondsforth...

Страница 13: ...igure 15 appear and the user can choose one of three options a Press to start the factory reset or b Press Cancel to return to the HELP screen or c Press Exit to abandon the factory reset entirely 7 A...

Страница 14: ...e phone Do not pair the phone Bluetooth to the SP 380P Open uPyxis Mobile App Ensure that the SP 380P is discoverable by following the steps in the Bluetooth Connection to Devices section Once the app...

Страница 15: ...well as update the firmware of the SP 380P by pressing Check Update If a firmware update is available press Get Firmware Once the new firmware is downloaded press Upgrade Firmware NOTE The firmware up...

Страница 16: ...Net Framework 4 5 if not previously installed on the PC the USB driver for the USB Bluetooth adapter MA NEB the USB RS485 adapter MA 485 and the main uPyxis Desktop application Double click the uPyxis...

Страница 17: ...Desktop App will default to the System screen From the System screen users can upgrade the firmware by selecting an appropriate firmware file contact service pyxis lab com for these firmware files and...

Страница 18: ...og List and selecting the desired datalog these are separated by month The SP 380P will then populate any relevant log event from the selected datalog which can be viewed in more detail by clicking Re...

Страница 19: ...hout any further key presses if it has not powered off If there are no key presses for 20 seconds the screen will darken and after another 20 seconds without key activity will power down The press of...

Страница 20: ...onfirm that the SP 380P is clean Repeat the process as needed until the Cleanliness Check shows Clean 11 3 Storage Do not expose the SP 380P to an extreme high or low temperature condition such as lea...

Отзывы: