USER GUIDE:
Ionic® Purification System
Purigen Biosystems
FOR RESEARCH USE ONLY.
Not for use in diagnostic procedures.
37
4.
Press
Yes
to confirm you understand that the cleaning procedure cannot be canceled after this point.
FIGURE 30:
Clean screen -
Step 2
5.
Pour 1 bottle of
Cleaning Solution A
into the large reservoir (
approximate volume 12 mL
).
FIGURE 31:
Clean screen -
Step 3
6.
Press the arrow on the right side of the system’s touchscreen to continue.
7.
The Ionic Purification System cover will close and complete the first step of the electrode cleaning process.
8.
After approximately 10 minutes, the first
Complete
screen is shown. Press the arrow on the right side of the system’s touchscreen
to open the instrument cover.