Cleaning the Ionic Purification System
The Ionic System Cleaning Kit is designed for use with the Ionic® Purification System. The kit contains fluidic chips and cleaning
solutions required to complete one system cleaning cycle.
NOTE
A system cleaning cycle
should be performed once every 30 protocol runs
. If the system is used less frequently, a cleaning
cycle should be performed every 90 days.
If more than 30 protocol runs have been completed without a cleaning, a warning will be displayed on the
Run Setup
screen.
FIGURE 27:
Run Setup screen: Instrument Requires Cleaning
The number of runs remaining until the next cleaning cycle is required can be found in the
Instrument Statistics
screen in the
on page 34 menu.
Ionic Purification System and associated consumables are intended for research use only. The table below lists the items included in
the Ionic System Cleaning Kit.
TABLE 16:
Cleaning Kit contents
Description
Quantity
Unit of Measure
Storage Requirements
Ionic System Cleaning Chip
2
Chip(s)
Room temperature
Cleaning Solution A
1
Bottle
Room temperature
Cleaning Solution B
1
Bottle
Room temperature
Cleaning Chip Seal
2
Seal(s)
Room temperature
In addition, you will need the following items:
TABLE 17:
User-supplied equipment
Name
P5000 Single Channel Pipette
P10000 Single Channel Pipette
Serologic
(5 or 10 mL)