FILTRATION & CHANGING FILTERS
ENSURE ALL FILTERS ARE FITTED CORRECTLY PRIOR TO USING THE MACHINE. DO NOT LIFT THE FILTERS USING THE HANDLES, THESE
ARE ONLY TO BE USED FOR SLIDING THE FILTERS OUT OF THE MACHINE. ENSURE THE AIRFLOW ARROW ON THE FILTER POINTS THE
CORRECT WAY.
Important note:
Always check the extraction nozzle, flexible hose, any pipework and the air inlet for debris build up before changing any filter.
This avoids false alerts related to airflow. Also, always change the pre-filter first (unless recently changed) and restart the
machine. If the alert continues, change the main filter.
Changing filters
Take care when removing main filters as they are heavy. Always wear the protective gloves and mask provided and dispose of
with the used filters in the bag provided. Protective eyewear should also be worn when changing filters.
Changing the pre-filter
The pre-filter is situated in the compartment underneath the main filter in Alpha & Xbase models. In FumeCube,
FumeCube
Lite, FumeSafe & FumeBuster models the prefilter pad is located above the main filter. In FumeCube MAX and
FumeCube PRO the Labyrinth pre-filter is located above the main filter.
Alpha, Alpha Dust & Xbase models: -
• Open the door
• Pull the filter (2) off the air inlet (1) and remove
• Insert the new filter, pushing the hole in the filter over the air inlet (1)
• Close and lock the door and restart the machine
FumeCube, FumeCube Lite, 2 Tip, FumeSafe & FumeBuster models: -
• Remove the lid
• Remove the old pre-filter pad and replace it with the new one. Make sure the harder, smoother side of the pad is facing the
filter. (On FumeSafe machines ensure the yellow side of the filter faces up)
• Replace the lid, secure the catches (or screws) and restart the machine
FumeCube MAX & FumeCube PRO models: -
• Remove the lid
• Pull the filter (2) off the air inlet (1) and remove
• Insert the new filter, pushing the hole in the filter over the air inlet (1)
•
Replace the lid, secure the catches (or screws) and restart the machine
Changing the main filter
Alpha & Xbase models: -
• Open the door
• Turn the filter locking handle (6) 180º anti-clockwise so that it is horizontal and in the filter unlocked position
• The filter (3) should drop slightly to allow release
• Remove the filter by sliding it towards you and remove it completely from the machine
• Insert the new filter (completely) into the machine ensuring that the airflow arrow on the label (7) is pointing up.
• (Purex recommend the pre-filter is also changed unless you know it has been already changed recently)
• Turn the filter locking handle 180º clockwise so that it is horizontal and in the filter locked position
• Close and lock the door and restart the machine
FumeCube, FumeCube MAX, FumeCube PRO, 2 Tip, FumeCube Lite, FumeSafe & FumeBuster models: -
• Remove the lid and lift the filter out of the machine
• Insert the new filter (completely) into the machine ensuring that the airflow arrow on the label (7) is pointing down.
• (Purex recommend the pre-filter is also changed unless you know it has been already changed recently)
• Replace the lid, secure the catches (or screws) and restart the machine