11.8
Can’t Create a Low Pressure Alarm
Possible Cause
Check
Corrective Action
Low Pressure Alarm set
point lower than default
setting of 0.15 PSI
Verify Low Pressure
Alarm set point
(Section 8.8.4 )
Adjust Low Pressure
Setting so that the
PRESSURE
reading
drops below verified set
point (Section 8.7.5 )
Defective Pressure
control
Verify that the Pressure
cannot be adjusted
(Section 8.7.5 )
Contact PUREGAS
Technical Support
(Section 11.16 )
Defective Dehydrator
Verify
PRESSURE
reading is lower than the
Low Pressure Alarm set
point (Sections 8.4.5.2
& 8.8.4 )
Contact PUREGAS
Technical Support
(Section 11.16 )
11.9 High Humidity
Possible Cause
Check
Corrective Action
High Humidity Alarm
set point too low
Verify High Humidity
Alarm set point
(Section 8.8.10 )
Raise Humidity Alarm
set point
(Section 8.8.10 )
Over 10% not
recommended
Defective Dehydrator
Humidity Alarm does
not clear within 8 hours
of first appearing
Contact PUREGAS
Technical Support
(Section 11.16 )
11.10 High Cabinet Temperature Alarm
Possible Cause
Check
Corrective Action
High Ambient
Temperature
Verify temperature of
Dehydrator operating
location. Recommended
ambient temperature is
40°-85°F.
Lower ambient
temperature of
Dehydrator operating
location
11.11 Low Cabinet Temperature Alarm
Possible Cause
Check
Corrective Action
Low Ambient
Temperature
Verify temperature of
Dehydrator operating
location. Recommended
ambient temperature is
40°-85°F.
Raise ambient
temperature of
Dehydrator operating
location