9.2.3
For P200WA model:
9.2.3.1
Disconnect the AC
Power Cord from the
Power Port on the back
of the Dehydrator.
9.2.3.2
Using a Voltmeter,
place the probes inside
the AC Power Cord
LINE
&
NEUTRAL
plug openings.
(Not GROUND)
The voltage should measure
85 - 264 VAC
.
9.2.3.3
Re-connect the AC Power Cord to the Power Port on the back of the
Dehydrator.
9.2.4
Power the Dehydrator
ON
(Section 8.3 ).