5.
Wipe the sensing electrode and the surrounding area with a dry paper tissue.
PAPER
TISSUE
SENSING
ELECTRODE
6.
Place a new silicone sheet (if used) and O-ring on the sensor. Make sure that the area is dry of
electrolyte.
O-RING
SILICON SHEET
Apply a drop of electrolyte on the sensing electrode.
SENSING
ELECTRODE
DROP OF
ELECTROLYTE
ENLARGED
VIEW
7.
Place a new membrane on the sensor, with the bead of electrolyte between the electrode and the
membrane.
ENLARGED
VIEW
MEMBRANE
White Membrane
P/N M-4400M
Caution: A GREEN tinted paper
separates each membrane.
DO NOT USE THIS PAPER
M-4400M Membrane
Color is White
Electrolyte
P/N 84004
NOTE: If reusing the old
O-ring and Silicon sheet, please
dry them thoroughly before
reinstalling.
CH
3
Br sensor cell
uses a White O-ring,
p/n 86001 and
Silicon sheet,
p/n 86002
40
NOTE: Never touch the center of the
membrane with bare fingers. Oil on
your fingers can block the
membrane
pores.