6/24 - pure STERIBASE 300
1025009-00
07/2020
General information
Electrical safety
– The device must not be connected to the mains socket
if the mains plug or cables are damaged. The mains
power cable must be replaced with a special power ca-
ble available from the manufacturer or its Customer
Service.
Order number: see page 12.
– This device is only to be operated with an alternating
voltage of 230-240 V with 50 Hz.
– The device must only be connected to a socket with
correctly installed earthing contacts.
– The system should be protected using a residual-cur-
rent circuit breaker (residual-current device) with a
maximum rated residual current of 30 mA.
– Before connecting the device, please make sure that
the power supply corresponds to the device's power
supply data.
Notes on installing and operating the device
– The device is intended for installation in rooms or facil-
ities that are dry and not exposed to splashing or drip-
ping water. It must not be installed in rooms where the
maximum relative humidity is over 60%.
– Adaptation of the device by the Customer Service de-
partment is required for operating altitudes greater
than 2000 m above sea level in order to guarantee
trouble-free operation.
– Trouble-free operation in mobile facilities (ship, bus,
train) is not guaranteed. If necessary, the prerequisites
can be created via modification by the Customer Ser-
vice.
– The installation location must be easy to access for
maintenance work.
– The device must be completely assembled and in-
stalled so that no UV-C radiation can escape.
– Please adhere to the prescribed maintenance inter-
vals.
– This device may be used by children over the age of 8,
persons with reduced physical, sensory or mental ca-
pabilities and persons who have little knowledge or ex-
perience of the device; however, they must be super-
vised at all times. Do not allow children to play with the
device. Cleaning and user maintenance tasks must not
be carried out by children.
– When the installation conditions are no longer fulfilled
(e.g. due to changes to the operating environment), the
device must not be operated any more.
– A socket with earthing contact must be provided by the
customer near to the device.
– The installation location must be selected so that de-
germed air can equally flow through the room.
– Avoid positioning in the immediate vicinity of seating
areas or workplaces as well as beds (hospital wards)
due to bothersome air flow.
– The air circulation should not be interrupted by barriers
such as room partitions, large plants, etc.
– When installing on a floor with carpet, more dust parti-
cles will be extracted that overload the filter additional-
ly. We recommend the operation on a smooth floor or
with a smooth subframe placed underneath (e.g a pan-
el).
– With longer periods of absence or during operating in-
terferences, pull out the mains plug to disconnect the
device from the power supply. Even when the mains
switch is located in the OFF position, the device is not
fully separated from the mains supply. We recommend
turning the device off at the mains plug if the device is
not used for temporary periods (e.g. overnight).
– The following minimum clearances must be observed
for the selection of the installation location.
a
≥
2 500 mm
b
≥
500 mm
h = 1640 mm
b
h
b
a
230 - 240 V
40325 / 1