PULSAR MEASUREMENT
17
Hazardous area location
Ex ia
(Intrinsically Safe) dBi-M transducers are certified for use in hazardous
zones 0, 1, 2 (gas) & 20, 21, 22 (dust).
The installation must comply with local regulations and the power supply
cable to the transducers shall meet the relevant requirements of clause 9 of
IEC60079-14:2013 / EN60079-14: 2014. The dBi-M series intelligent
transducer has bare wires; ensure that suitably specified barriers are
employed.
Marking for the dBi-M transducer:
Zener Barriers
RS485 -
RS485 +
Power
0V/Gnd
Red
White
Orange
Black
Zener barriers
Power:
Uo max. = 15.1V, R.min = 44.6R
Signal:
RS485 -, RS485 +
Uo max. = 10.3V, R.min = 49R
Safe Area
IS earth
IS earth
IS earth
Screen
NC
Entity parameters:
Power:
Ui = 15.1V
Li = 350mA
Pi = 0.65W
Signal:
Ui = 10.3V
Li = 350mA
Pi = 0.65W
Uo = 6.51V
Lo = 132mA
dBi-M
transducer
Hazardous area
Zones 0, 1, 2 (gas) & 20, 21, 22 (dust)