P809 Fail-safe Time
In the event of a fail-safe condition occurring the failsafe timer determines
the time before fail-safe mode is activated.
Default = 2.00 mins
If the timer activates, the unit goes into
fail-safe
, as determined by
P808
,
(
Display
and
P840
(
mA Output
). When this happens, you will see the
message “
LOE
!” on the display.
When a valid measurement is obtained then the display, relays and mA
output will be restored, and the timer is reset.
Compensation
P851 Measurement Offset
The value of this parameter is added to the measured distance, in
Measurement Units
(
P104
).
This Offset will be added to the level, as derived from the transducer, and
will affect everything including the reading on the display, the relay
setpoints and the mA output.
P852 Temperature Source
This parameter determines the source of the temperature measurement. By
default,
it is set to internal (transducer) (
P852=1
), which will automatically
detect the temperature from the transducer. If for any reason, no
temperature input is received, then the
Fixed Temp
value is used, as set by
P854
.
The temperature source can be specifically set as follows:
OPTION
DESCRIPTION
1 = Internal (Default)
Always uses temperature reading from
transducer.
3 = Fixed
Always uses fixed temperature (P854)
P854 Fixed Temperature
This parameter sets the temperature, in degrees centigrade to be used if
P852
(
Temperature Source
) =
3
.
Default = 20
o
C
Содержание dBi 10