background image

EN 

Pulse Oximeter tester) could measure the available pulse wave with the amplitude of 6% of the 
simulation pulse wave amplitude.  

11.

 

Interference Resistance Capacity against ambient light: Device works normally when BIO-TEK INDEX 
Oximeter tester exerts interfering signal test.  

Classification 

1.

 

Management Class for Medical Devices: II Equipment  

2.

 

Anti-electric Shock Type: Internally powered equipment 

3.

 

Anti-electric Shock Degree: Type BF equipment 

Maintenance and Preservation 

1.

 

Charge batteries timely when the battery power is too low.  

2.

 

Clean the surface of fingertip oximeter before it is used to diagnose patients.  

3.

 

It is recommended to preserve the product in -10 - 40°C (14 - 104°F) and humidity is 10%-80%.  

4.

 

It is recommended that the product should be kept dry any time. A wet ambience might affect its 
lifetime and even damage the product. 

Product Accessories  

1.

 

User manual 

2.

 

Charging cable 

3.

 

Adapter (optional)  

Guidance and Manufacturer’s Declaration

 - Electromagnetic Radiation for Other Equipments 

and Systems 

The Pulse Oximeter is designed to be used in specified electromagnetic environment. Users of the 
Pulse Oximeter must use it in the following environments.   
Radiation Test  

Compliance  

Electromagnetic environment 
guidance  

RF-interference CISPR 11  

Group 1  

RF signal of Pulse Oximeter is 
simply created by its internal 
function. Therefore, its RF 
interference is very low and is 
not likely to cause any 
interference to nearby 
electronic equipment. 

RF-interference CISPR 11 

Class B  

The Pulse Oximeter applies to 
all establishments, including 
domestic establishments and 
those directly connected to the 
public low-voltage power supply 
network that supplies buildings 
used for domestic purposes. 

Troubleshooting 

Problem 

Possible Reason 

Solution 

The SpO2 and Pulse Rate can 
not be shown normally 

1. Finger is not properly 
positioned 
2. P

atient’s SpO2 is too low to 

be measured. 

1. Place the finger properly and 
try again.  
2. Try again; If you are sure the 
device works all right, go to a 
hospital timely for exact 
diagnosis. 

The SpO2 and Pulse Rate are 
shown unsteadily 

1. Finger is not placed inside 
deep enough. 
2. Finger is trembling or the 
patient is moving. 

1. Place the finger properly and 
try again.  
2. Please remain at rest. 

The device can not be turned 
on 

1. Inadequate power or power 
off.  
2. Oximeter might be damaged. 

1. Charge battery timely. 
3. Please contact the local 
customer service. 

Display is suddenly off  

1. The device automatically 

1. Normal. 

Содержание PO-210B

Страница 1: ...OX PO 210B BEDIENUNGSANLEITUNG Pulse Oximeter PO 210B Instruction manual Pulsiox metro PO 210B Manual de instrucciones Oxym tre de Pouls PO 210B Mode d emploi Pulsossimetro PO 210B Instruzioni per l u...

Страница 2: ...feld von gr ter Wichtigkeit Das Pulsoximeter ist klein und tragbar hat einen geringen Stromverbrauch und eine komfortable Bedienung Messprinzip Das Prinzip des Pulsoximeters Es wird eine mathematische...

Страница 3: ...n Verordnungen und Recyclinghinweise zur Entsorgung bzw zum Recycling des Ger tes und der Ger tekomponenten Anwendungsbereich Verwendungszweck Das Pulsoximeter ist f r Kinder unter 12 Jahren geeignet...

Страница 4: ...n neben Restore angezeigt wird dr cken Sie die Taste lange 3s und alle Einstellungen werden wieder auf die Werkseinstellungen zur ckgesetzt Achtung Bei Verwendung des Ger ts im Freien oder unter stark...

Страница 5: ...her Schock Grad Typ BF Wartung und Instandhaltung 1 Laden Sie den Akku rechtzeitig auf wenn der Akkustand zu niedrig ist 2 Reinigen Sie die Oberfl che des Oximeters bevor Sie es zur Messung am Patient...

Страница 6: ...Laden Sie den Akku auf Garantie und Service Wir leisten 1 Jahre Garantie f r Material und Fabrikationsfehler des Produktes Die Garantie gilt nicht im Falle von Sch den die auf unsachgem er Bedienung b...

Страница 7: ...Vor direkter Sonne sch tzen Die CE Kennzeichnung bescheinigt die Konformit t mit den grundlegenden Anforderungen der Richtlinie 93 42 EEC f r Medizinprodukte Ein Aus Technische nderungen vorbehalten P...

Страница 8: ...s is established by exerting Lambert Beer Law according to Spectrum Absorption Characteristics of Reductive Hemoglobin R Hb and Oxyhemoglobin O2 Hb in glow and near infrared zones Operation principle...

Страница 9: ...and SpO2 sensor Features OLED Display Product adopts double color OLED display can show the six different display mode Low power consumption continuously work for more than six hours with one recharg...

Страница 10: ...item then press button 3s to setup data Choose or to plus or minus values Menu 1 Menu 2 Detailed Description of Product Functions 1 Display type OLED Display 2 SpO2 Measurement range 70 99 Accuracy 2...

Страница 11: ...er must use it in the following environments Radiation Test Compliance Electromagnetic environment guidance RF interference CISPR 11 Group 1 RF signal of Pulse Oximeter is simply created by its intern...

Страница 12: ...d part European Representative WEEE 2002 96 EC Serial number Refer to instruction manual Lot number Caution Date of manufacture IP22 International Protection degree Manufacturee Humidity range Tempera...

Страница 13: ...medici n del pulsiox metro funciona de la siguiente manera Se establece una f rmula de experiencia de procesamiento de datos aplicando la Ley de Lambert Beer seg n las caracter sticas de absorci n del...

Страница 14: ...utilizar para medir la saturaci n y la frecuencia card aca de los humanos a trav s del dedo Se puede utilizar en hospitales familias escuelas y centros m dicos NOTA 1 Es posible que detecte que las im...

Страница 15: ...zonas con mucha luz ajuste el brillo de la pantalla adecuadamente Seleccione un menor brillo para ahorrar bater a 2 Men 2 Cuando aparezca el indicador en el men Sounds Setup mantenga pulsado el bot n...

Страница 16: ...da arlo Accesorios del producto 1 Manual de usuario 2 Cable de carga 3 Adaptador opcional Gu a y declaraci n de fabricaci n Radiaci n electromagn tica para otros equipos y sistemas El pulsiox metro es...

Страница 17: ...u s de 8 segundos 2 El dispositivo tiene poca bater a 1 Normal 2 Cargue la bater a S mbolos y Definiciones Symbol Description Symbol Description Protecci n y aislamiento del dispositivo el ctrico Repr...

Страница 18: ...n et r p tabilit Principe de mesure Le principe de mesure de l oxym tre fonctionne de la mani re suivant il tablit une formule d exp rience de traitement de donn es en appliquant la loi de Lambert Bee...

Страница 19: ...on Pr vue L oxym tre de doigt peut tre utilis pour mesurer la saturation et la fr quence cardiaque des humains travers le doigt Il peut tre utilis dans les h pitaux en famille dan les coles ou les cen...

Страница 20: ...nt plus de trois secondes tous les r glages retourneront la configuration d usine Attention Si le dispositif va tre utilis l ext rieur ou dans des zones avec beaucoup de lumi re ajustez la brillance d...

Страница 21: ...hargez la batterie que lorsque le niveau de la batterie est faible 2 Nettoyez la surface du bout du doigt avant de l utiliser 3 Conservation du produit 10 C 40 C 14 104 F et humidit de 10 80 4 Il est...

Страница 22: ...2 L oxym tre pourrait tre endommag 1 Chargez la batterie 2 Veuillez contacter votre distributeur Les indicateurs s teignent subitement 1 Le dispositif s teint automatiquement s il ne d tecte pas de s...

Страница 23: ...segue Viene stabilita una formula matematica che si avvale della Legge di Lambert Beer in conformit con le caratteristiche dello spettro di assorbimento dell emoglobina riduttiva R Hb e dell ossiemogl...

Страница 24: ...er misurare la saturazione e la frequenza cardiaca nelle persone e si pu utilizzare in ospedali casa scuole e centri medici NOTA 1 possibile che le immagini presenti in queste istruzioni siano differe...

Страница 25: ...lta luce regolare la luminosit del display di conseguenza 2 Men 2 quando appare l indicatore sul men Sounds Setup premere il pulsante e tenerlo premuto per oltre tre secondi per accedere al secondo me...

Страница 26: ...i fabricazione radiazione elettromagnetica per altri apparecchi e sistemi Questo pulsossimetro progettato per essere utilizzato in un contesto elettromagnetico specifico Gli utenti devono utilizzarlo...

Страница 27: ...2 Ricaricare la batteria Simboli e Definizioni Simbolo Descrizione Simbolo Descrizione Protezione e isolamento del dispositivo elettrico Rappresentante dell Unione Europea Raccolta differenziata WEEE...

Страница 28: ...ne question s il vous pla t contacter Se si dispone di una domanda si prega di contattare Novidion GmbH Fuggerstr 30 51149 K ln Tel 49 2203 9885 200 Fax 49 2203 9885 206 info novidion de www pulox de...

Отзывы: