Intended Use

 

 

D616.131.00/10 en 

 

Intended Use 

DIADOS E  is  a  diagnostic  dosemeter  according  to 

IEC 61674 and can be used for acceptance testing 

and  constancy  checks.  It  can  be  employed  in  con-

ventional  X-ray  diagnostics,  in  dental  and  mammo-

graphic  applications  as  well  as  with  CT  systems. 

DIADOS E  is capable  of measuring dose  and  dose 

rate  in  diagnostic  X-ray  applications,  dose  length 

product  and  the  dose  length  product  rate  in  CT 

applications and current and charge if used as elec-

trometer.  DIADOS E  is  capable  of  measuring  in 

front of and behind phantoms. 

The  following  detectors  are  required  for  the  meas-

urement: 

  DIADOS  Diagnostic  Detector  T60004  for  con-

ventional diagnostic radiology and for dental im-

aging  

  DIADOS MAM Detector T60005 for mammogra-

phy  

  CT  ion  chamber  3.14 cm³  TM30009  with  CT 

adapter T16018 

  CT ion chamber 300 mm long TM30017 with CT 

adapter T16018 

 

Note: 

In  the  United  States,  federal  law  restricts 

this device to sale by or on the order of a 

physician. 

 

NOTE 

Please observe the user manuals of all connect-

ed detectors! 

 

Safety Information  

 

This is the safety  alert symbol. It  is used to 

alert the user to potential hazards. Obey all 

safety  messages  that  follow  this  symbol  to 

avoid  possible  bodily  injury  or  equipment 

damage.  

All  safety  messages  consist  of  the  following  com-

ponents: 

 

–  Safety alert symbol and signal word   
–  Type of danger 

 

–  Source of danger 

 

–  Consequence 

 

–  Measures to prevent hazard 

Signal Words 

DANGER 

Indicates an imminent hazard. If not avoided, the 

hazard will result in death or serious injury. 

 

WARNING 

Indicates  a  hazard.  If  not  avoided,  the  hazard 

can result in death or serious injury. 

 

CAUTION 

Indicates  a  potential  hazard.  If  not  avoided,  the 

hazard  could  result  in  minor  injury  or  product  / 

property damage. 

 

NOTE 

Provides  useful  information  to  assure  that  you 

get the most from your equipment. 

 

Содержание DIADOS E

Страница 1: ...User Manual DIADOS E T11035 firmware 1 30 or higher Contents Operating Manual Technical Manual Service Manual D616 131 00 10 en 2015 10 Hn Az...

Страница 2: ...ite for proper product performance and correct operation Operator safety specified measuring accuracy and interference free operation can be guaran teed only if original products and parts are used Fu...

Страница 3: ...concern ing the products of interest No part of the technical documentation may be reproduced without written permission from PTW Freiburg PTW Freiburg is registered manufacturer ac cording to the El...

Страница 4: ...ing the Detector 20 2 4 Positioning DIADOS E 21 2 5 Basic Settings Setup Menu 22 2 6 Modes of Operation 23 2 7 Measuring Quantities 23 2 8 Measuring Ranges 25 2 9 Digital Resolution 27 2 10 Entering t...

Страница 5: ...ndustry Standard Compliance 49 Index 51 Illustrations Figure 1 Definition of patient environment 10 Figure 2 DIADOS E connection Panel 12 Figure 3 DIADOS E display 12 Figure 4 DIADOS E keypad 13 Figur...

Страница 6: ...apter T16018 CT ion chamber 300 mm long TM30017 with CT adapter T16018 Note In the United States federal law restricts this device to sale by or on the order of a physician NOTE Please observe the use...

Страница 7: ...placed immediately When disconnecting the device from the pow er line first remove the plug from the wall out let then disconnect the power cord from the device WARNING Devices on which moisture conde...

Страница 8: ...ble of applying it properly The operator must be trained in the use of the device WARNING Magnetic and electrical fields are capable of in terfering with the proper performance of the de vice Equipmen...

Страница 9: ...atings indicated on the name plate of the device Set up the device in a location which affords sufficient ventilation Always observe the ambient conditions as indi cated in the Technical Specification...

Страница 10: ...to the section Technical Specifications Identification mark of an officially calibrated device according to the German Metrology and Verification Act Mess und Eichverordnung MessEV The device is prov...

Страница 11: ...ti range power sup ply or from rechargeable batteries For officially calibrated dosemeters On the side there is an official seal If the official seal is damaged or removed the official calibration wil...

Страница 12: ...different readouts during a measurement Figure 3 DIADOS E display 1 Resolution greater than 0 5 greater than 1 less than 1 2 Unit C Gy or R displayed in reverse video in pulsed mode 3 Application RAD...

Страница 13: ...al mode in the pulsed mode the unit is displayed in reverse video RANGE toggles between the measuring ranges Auto Range Low Medium and High combination of keys ALT MODE RESET resets the display to zer...

Страница 14: ...RESET key A measurement initiated with Autostart can also be aborted with RESET All parameter readings will be reset to zero 1 5 4 Zero Adjustment ZERO The keys ALT RESET will cause DIADOS E to perfor...

Страница 15: ...ctive only when a radi ological unit is selected Setup menu Two calibra tion factors can be entered for each application one for application with additional filtration Att and one for application with...

Страница 16: ...ssible 1 5 8 Selecting the Measuring Range RANGE The measuring range is selected with the keys ALT MODE a Low Low measuring range suitable for fluoroscopy and measurements behind the phantom b Medium...

Страница 17: ...range is selected after the measurement After the end of the measurement and when the dose is displayed the measuring range remains set until RESET is activated 1 5 9 ALT Key In combination with the...

Страница 18: ...n each case When radiation is removed DIADOS E will display the dose rate value measured 3 s before removal of radiation b Determining the Average Dose Rate Over a Period of 20 s The timer interval sh...

Страница 19: ...ome in contact with electrolyte with water In case of eye contact immediately rinse eyes with water and consult doctor Do not open rechargeable batteries WARNING Improper handling of batteries Risk of...

Страница 20: ...nected from mains NOTE In CT applications CT chamber and CT junc tion box operate DIADOS E only on battery power refer to section 2 3 2 Press the ON OFF key to turn DIADOS E on The ON OFF key must be...

Страница 21: ...ger the device starting a measurement This phenomenon does not occur when the device is exclusively powered from batteries In CT applications CT chamber and CT junction box operate DIADOS E only on ba...

Страница 22: ...it You can choose among the units Gy R and C As soon as the DIADOS E is officially calibrated the unit R can not be selected Timer For measurements in normal mode the timer can be set to OFF 20 s or 2...

Страница 23: ...tomatic recognizing of the signal end Then the measurement is regarded as completed and will be finished Times longer than 1 s are meaningful if radiation distances are bigger than 1 s to expect The t...

Страница 24: ...well as Mo Anode 25 m Rh 2 mm Al W Anode 50 m Rh 2 mm Al and Rh Anode 25 m Rh 2 mm Al W Anode 50 m Ag 2 mm Al W Anode 0 7 mm Al 2 mm Al in range 25 to 45 kV dose or dose rate exposure time or irradiat...

Страница 25: ...Gy cm s 0 02 Gy cm s 17 Gy cm s 2 Gy cm s 1700 Gy cm s Current 3 pA 2 nA 0 3 nA 200 nA 0 03 A 20 A Table 1 Range limits for dose rate measurement Typical values for a minimum resolution of 1 in depend...

Страница 26: ...ate an elevated measuring error due to insufficient resolution f Hz Tpulse ms Low nGy pulse Medium Gy pulse High Gy pulse 150 3 3 2 0 1 1 9 5 5 4 0 12 0 07 0 07 11 0 6 0 3 5 100 5 3 0 1 1 10 6 3 5 0 1...

Страница 27: ...resolution for the charge and current readout is 30 fC and 30 fA respectively 2 10 Entering the Chamber Calibration Factor If DIADOS E is not officially calibrated the stored calibration data can be...

Страница 28: ...dose dose per pulse and pulse Select measuring range ALT RANGE LOW MEDIUM HIGH or Auto Range Display changeover ALT In normal mode toggle between dose dose rate and time or in pulsed mode toggle betwe...

Страница 29: ...FILTER Select measuring range with RANGE If the range changes automatically during the measurement the readout changes to AUTO M or AUTO H Select operating mode with MODE Perform a zero adjustment key...

Страница 30: ...icult to remove Keep the X ray path clear of adhesive tape 4 2 3 Starting the Measurement Autostart DIADOS E is permanently in the Autostart mode This means that a dose measurement begins as soon as a...

Страница 31: ...ement initiated with Autostart ends auto matically when no measuring signal is detected for more than the measuring delay time It is not possi ble to stop the measurement manually RESET aborts a measu...

Страница 32: ...cation of application and filter Figure 15 Display in electrometer mode Use the RANGE function to select the measuring range LOW MEDIUM or HIGH Table 1 section 2 8 shows the upper range limits for the...

Страница 33: ...ution is insufficient one or two asterisks or are displayed to the left of the measured value If the digital resolution of the display is below 1 two asterisks will appear with a resolution between 0...

Страница 34: ...lect the operating mode and the necessary filter setting Typical settings RQT9 0 25 mm Cu for measurements free in air and for measurements in phantoms The LOW range is suitable for most measurements...

Страница 35: ...etector housing marks the active field You can install the detector in the X ray path using Velcro tape The Velcro tape should be attached on the back of the detector for this purpose CAUTION Improper...

Страница 36: ...ect the appropriate radiation quality with the FILTER key at the DIADOS E see section 1 5 6 CAUTION Improper handling Equipment damage Ensure that the cable is routed so as not to im pede the range of...

Страница 37: ...corrupted RESET continue with default values Erroneous device settings problem is not severe Confirm message with RESET and continue Values for range mode filter etc are reset to the defaults Radiolo...

Страница 38: ...User Manual RS232 Interface for PTW DIADOS E LCD does not illumi nate The rechargeable batteries are depleted Connect the multi range power supply The bat teries charge as soon as the power supply is...

Страница 39: ...z of March 23 1992 BGBl I S 711 in the presently relevant version approval sign 23 04 03 01 issued to PTW Freiburg Use Diagnostic dosemeter acc to IEC 61674 for dose and dose rate measurement in diagn...

Страница 40: ...0 01 9999 s in the range Low Digital resolution DIADOS E electrometer 30 fA or 30 fC DIADOS Diagnostic Detector T60004 1 nGy s or 1 nGy DIADOS MAM Detector T60005 2 nGy s or 2 nGy Leakage current zer...

Страница 41: ...mm x 44 mm x 200 mm incl socket Weight 500 g incl batteries The product fulfills the following requirements of EN55011 IEC CISPR11 Emissions test Compliance Electromagnetic environment guidance RF em...

Страница 42: ...DIADOS E T11035 Technical Manual Technical Specifications 42 D616 131 00 10 en...

Страница 43: ...e from the power line Wipe the device clean with a moist cloth Do not let liquid enter the device Do not apply spray to the device Use a mild soap solution to clean the device Due to the device geomet...

Страница 44: ...Repairs may only be carried out by PTW Freiburg or by persons companies authorized by PTW Freiburg 8 3 Safety Inspections Only regularly inspected products are fail safe To preserve the functional se...

Страница 45: ...re under the official cali bration procedure if it has been officially re calibrated 8 6 Calibration When using not officially calibrated dosemeters we recommend having the system calibrated every 2 y...

Страница 46: ...harged again approximately every 3 months Replacing the batteries CAUTION Improper handling of batteries Equipment damage Do not operate the DIADOS E with disposable batteries Use only the special rec...

Страница 47: ...t be disposed of in compliance with the applicable waste control regulations The different materials must be separated and re cycled as appropriate DIADOS E does not contain hazardous materials The el...

Страница 48: ...ical equipment Dosimeters with ionization chambers and or semi conductor detectors as used in X ray diagnostic imaging 6 IEC 61010 1 Safety requirements for electrical equipment for measure ment contr...

Страница 49: ...c or hazardous substances or elements contained in electronic information products will not leak or mutate under normal operating conditions so that the use of such electronic information products wil...

Страница 50: ...is below the limit requirement in SJ T11363 2006 X Indicates that this toxic or hazardous substance contained in at least one of the homogeneous materials used for this part is above the limit requir...

Страница 51: ...3 L Literature 48 M Maintenance 44 Measurement 29 Measurement at dental X ray units 35 Measurement start stop 14 Measurements at panorama X ray units 36 Measurements in Computed Tomography 21 34 Measu...

Отзывы: