DIADOS E
T11035
Operating Manual
– Putting the System into Operation
D616.131.00/10 en
27
2.9 Digital Resolution
The minimum digital resolution of the dose and
dose rate readout within the measurement range
limits shown in tables 1 and 2 is 1%. The display
range for radiography and fluoroscopy time is
(0.001 … 9999) s, in the LOW range (0.01 ...
9999) s.
The digital resolution for the charge and current
readout is 30 fC and 30 fA, respectively.
2.10 Entering the Chamber
Calibration Factor
If DIADOS E is
not
officially calibrated, the stored
calibration data can be completed or corrected (e.g.
for a new detector) by means of a command sent
via the RS232 interface. To officially calibrate
DIADOS E for specific applications, the device and
all detectors to be used should be shipped to PTW
Freiburg.
For information on the command, please refer to
the separate interface manual
„RS232 Interface for
PTW-DIADOS E
“ D616.131.01.