ADG400 User Manual
18
99998 Issue 1, November 2022
OPERATION
3.6.10 Saturator Filling Mode
It is necessary to use the saturator filling mode when adding water to the saturators.
Item
Description
Start
Begin de-pressurization of saturators and water traps to allow them to
be filled.
Stop
Begin re-pressurization of saturators and water traps after filling.
Back
Returns to the settings screen
Figure 14
Saturator Filling Screen
To fill the saturators, follow the on-screen instructions. Both saturators must be filled
for proper operation, although water is consumed by the first saturator at a faster rate.
The water level in both the saturator and its associated water trap should be considered
as per the below diagram.
Figure 15
Saturator Fill Levels