ADG400 User Manual
6
99998 Issue 1, November 2022
OPERATION
A typical calibration to -80 °Cfp, for the Michell Instruments Easidew, would use the
following timings:
Set Point
(°C Dew/Frost Point)
Minimum Stabilization Time
DRYER
48 hours
-80
12 hours
-70
8 hours
-60
4 hours
-50
2 hours
-40
1 hour
-30
1 hour
-20
1 hour
-10
1 hour
0
1 hour
+10
1 hour
+20
1 hour
Table 1
Minimum Stabilization Times
3.3
HMI
The ADG uses a 3.5" full-color touch-screen controller.
3.4
Screensaver
When idle for 10 minutes, the display of the ADG will be set to screensaver mode, which
dims the display. This helps to maintain the brightness of the backlight as the unit is
constantly powered on.
The length of time before the screensaver is active is configurable in the settings menu.