56
Protocol Systems, Inc.
Respond to An Equipment Alert
When the monitor detects an equipment problem, it produces an audible alert tone every five seconds.
It also displays an equipment alert message similar to the following
:
1.
Press any key at the bottom of the screen to silence the alert tone (or press the
Suspend/Resume
Alarms
key in the upper right corner to silence the alert tone for 90 seconds).
2.
Determine what caused the problem and correct it. For descriptions of equipment alert
messages and suggested responses, see page 57.
Sometimes an equipment alert also causes one or more patient alarms. Patient alarms have a higher
priority than equipment alerts and are displayed first. Respond to the patient alarm or alarms (see
page 53), then respond to the equipment alert.
If you turn off any alarm limits while responding to a patient alarm, be sure to restore the appropriate
alarm limits before resuming patient monitoring.
NET
OFF
HR
MON
17:05:10
PRINT FAULT
II
BR
P
1
( 96)
BPM
mmHg
(M)
TEMP
100.4
98.6
1.8
97
121
79
S
D
P
2
( 15)
mmHg
(M)
25
9
S
D
%
°
F
MCO
2
S
p
O
2
mmHg
38
Br/m
12
ADULT
ON NETWORK
T1
T2
∆
T
mV
cm
1
112
EQUIPMENT ALERT
NIBP FAULT
CUFF NOT DETECTED
PRESS ANY KEY BELOW TO ACKNOWLEDGE
ALLEN KIM
General
equipment alert
message
Specific problem
Press any key to
silence the alert tone.
Содержание Propaq 242
Страница 6: ...6 Protocol Systems Inc...
Страница 18: ...18 Protocol Systems Inc...
Страница 26: ...26 Protocol Systems Inc...
Страница 50: ...50 Protocol Systems Inc...
Страница 52: ...52 Protocol Systems Inc...
Страница 64: ...64 Protocol Systems Inc...
Страница 66: ...66 Protocol Systems Inc...
Страница 72: ...72 Protocol Systems Inc...
Страница 74: ...74 Protocol Systems Inc...
Страница 78: ...78 Protocol Systems Inc...
Страница 84: ...84 Protocol Systems Inc...
Страница 86: ...86 Protocol Systems Inc...
Страница 94: ...94 Protocol Systems Inc...
Страница 120: ...120 Protocol Systems Inc...